FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 7337001 · Received March 14, 2018

Report

Report Number
1645337-2018-01409
Event Type
Injury
Date Received
March 14, 2018
Date of Event
February 13, 2018
Report Date
February 13, 2018
Manufacturer
MENTOR TEXAS
Product Code
FTR
PMA / PMN Number
P030053
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION WAS RECEIVED WITH THE FIELD EXPERIENCE REPORT ON (B)(6) 2018. THE IMPLANT DATE WAS UPDATED TO (B)(6) 1999. THE EXPLANT DATE WAS UPDATED TO (B)(6) 2018. THE BRAND NAME, CATALOG NUMBER, LOT NUMBER, AND SERIAL NUMBERS WERE OBTAINED. THE IMPACTED PRODUCT WAS A 300CC MENTOR MEMORYGEL BREAST IMPLANT. THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION ON (B)(6) 2018. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATION AND ANALYSIS HAVE BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. A DEVICE HISTORY RECORD REVIEW IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 300CC MENTOR MEMORYGEL BREAST IMPLANT, LOT #190272, SERIAL # (B)(4), CATALOG #3543007 MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

THE DEVICE HISTORY RECORD (DHR) FOR THE LOT NUMBER 186548 WAS REVIEWED ON (B)(6) 2018 AND IT WAS VERIFIED THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATIONS AND PROCEDURES. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

THE EVALUATION OF THE RETURNED DEVICE WAS COMPLETED BY THE FAILURE ANALYSIS LAB ON (B)(6)2018. DEVICE EVALUATION SUMMARY: IT WAS REPORTED THAT A 50-YEAR-OLD FEMALE UNDERWENT PRIMARY BREAST AUGMENTATION WITH AN UNKNOWN MENTOR SALINE PROSTHESIS AND EXPERIENCED RUPTURE POST PROCEDURE. UPON RECEIPT TO MENTOR, THE DEVICE CONTAINED NO GEL. YELLOW MATERIAL WAS OBSERVED WITHIN THE DEVICE. GEL WAS OBSERVED ON THE SHELL SURFACE. VISUAL EVALUATION OF THE DEVICE REVEALED A RENT WITHIN AN AREA OF SILTEX CRACKING ON THE POSTERIOR VIEW MEASURING 1.6 CM. NO OTHER ANOMALIES WERE FOUND. MENTOR PRODUCTS ARE 100% VISUALLY INSPECTED PRIOR TO RELEASE IN ADDITION TO THOROUGH IN-PROCESS TESTING DURING SEVERAL STAGES OF THE MANUFACTURING PROCESS. THE MENTOR PRODUCT ANALYSIS LAB CONCLUDED THAT THE RENT AND SILTEX CRACKING OCCURRED SOMETIME SUBSEQUENT TO THE REMOVAL OF THE DEVICE FROM ITS PROTECTIVE PACKAGING. THE COMPLAINT OF RUPTURE WAS CONFIRMED SINCE A RENT WAS FOUND ON THE DEVICE. HOWEVER, AT THIS POINT IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF SUCH DAMAGE OR THE ORIGIN OF THE YELLOW MATERIAL OBSERVED. SILTEX CRACKING IS DEFINED AS A MATERIAL SEPARATION OCCURRING IN THE SILTEX LAYER AND CAN EVENTUALLY PROGRESS THROUGH THE SHELL OF SILTEX DEVICES. MENTOR BELIEVES THAT SILTEX CRACKING IS ULTIMATELY A RESULT OF STRESSES CAUSED BY FOLDING OF THE SHELL IN-VIVO. RUPTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET; HOWEVER, COMPLAINT TRENDS FOR MENTOR DEVICES WILL CONTINUE TO BE MONITORED BY QUALITY ASSURANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. IT IS UNKNOWN AT THIS TIME IF THE DEVICE WILL BE MADE AVAILABLE FOR RETURN. AS A RESULT, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO DEVICE HISTORY RECORD (DHR) REVIEW COULD BE PERFORMED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) FEMALE UNDERWENT PRIMARY BREAST AUGMENTATION WITH AN UNKNOWN MENTOR SALINE PROSTHESIS AND EXPERIENCED RUPTURE POST PROCEDURE. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
183232 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 186548

Patients

Seq Age Sex Outcome Treatment
1 50 YR Other| R