MENTOR SILTEX ROUND MODERATE PROFILE
Report
- Report Number
- 1645337-2018-01441
- Event Type
- Injury
- Date Received
- March 13, 2018
- Date of Event
- January 3, 2018
- Report Date
- July 13, 2018
- Manufacturer
- MENTOR TEXAS
- Product Code
- FWM
- UDI-DI
- 00081317001775
- PMA / PMN Number
- P990075
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
ON 06/26/18, IT WAS REPORTED TO MENTOR THAT THE LOT NUMBER HAS BEEN PROVIDED WAS MANUFACTURED IN (B)(4). ON 07/09/18, IT WAS REPORTED TO MENTOR THAT THE DEVICE HISTORY RECORD ( DHR) OF LOT NUMBER 211868 REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
ON 07/13/2018, IT WAS REPORTED TO MENTOR THAT THE DEVICE EVALUATION AND INVESTIGATION HAVE BEEN COMPLETED. UPON RECEIPT BY MENTOR, TWO DEVICES WERE RETURNED. THE DEVICE A RETURNED WITHOUT FLUID. NO FOREIGN MATERIAL WAS OBSERVED WITHIN THE DEVICE OR ON THE SHELL SURFACE. PRODUCT EVALUATION TEAM DISCOVERED A RENT MEASURING APPROXIMATELY 1 CM WITHIN AN AREA OF SILTEX CRACKING ON THE POSTERIOR ASPECT. NO OTHER ANOMALIES WERE DISCOVERED. OTHERWISE, ON THE DEVICE B, IT RETURNED WITHOUT FLUID. NO FOREIGN MATERIAL WAS OBSERVED WITHIN THE DEVICE OR ON THE SHELL SURFACE. PRODUCT EVALUATION TEAM DISCOVERED A RENT MEASURING APPROXIMATELY 1.2 CM AND RENT B MEASURING APPROXIMATELY 0.7 CM; ALL LOCATED ON THE ANTERIOR ASPECT. NO OTHER ANOMALIES WERE DISCOVERED. THE COMPLAINT WAS CONFIRMED SINCE A RUPTURE WAS FOUND ON THE DEVICE. HOWEVER, AT THIS POINT IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF SUCH DAMAGE. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. SILTEX CRACKING IS DEFINED AS A MATERIAL SEPARATION OCCURRING IN THE SILTEX LAYER AND CAN EVENTUALLY PROGRESS THROUGH THE SHELL OF SILTEX DEVICES. MENTOR BELIEVES THAT SILTEX CRACKING IS ULTIMATELY A RESULT OF HIGH STRESSES CAUSED BY ACUTE FOLDS (I.E., SEVERE ¿V¿ OR TIGHT FOLDS) IN THE RTV SHELL OF SILTEX BREAST IMPLANTS. THE SEVERITY LEVEL FOR THIS COMPLAINT CODE WAS DETERMINED TO BE A S3, WHICH IS DEFINED AS: NECESSITATES INTRUSIVE MEDICAL OR SURGICAL INTERVENTION. BECAUSE PRODUCT EVALUATION TEAM WAS UNABLE TO CONFIRM ANY UNUSUAL FAILURE MODE, NO CORRECTIVE ACTION WILL BE TAKEN AT THIS TIME. MANUFACTURER¿S REFERENCE NUMBER (B)(4).
ON (B)(6) 2018, IT WAS REPORTED THAT MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO DEVICE HISTORY RECORD (DHR) REVIEW COULD BE PERFORMED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4)
IT WAS REPORTED THAT A (B)(6) -YEAR-OLD FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH A SALINE MENTOR SILTEX ROUND MODERATE PROFILE 275CC AND EXPERIENCED DEFLATION ON THE LEFT BREAST IMPLANT AFTER IMPLANTATION. AS A RESULT, THE PATIENT HAD UNDERGONE EXPLANTATION AND REPLACEMENT ON (B)(6) 2018.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 179828 | MENTOR SILTEX ROUND MODERATE PROFILE | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | MENTOR TEXAS | 00081317001775 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR | Required Intervention |