A1A-360
Report
- Report Number
- 8031673-2018-02808
- Event Type
- Malfunction
- Date Received
- March 13, 2018
- Date of Event
- January 13, 2017
- Report Date
- March 13, 2018
- Manufacturer
- TOSOH CORPORATION
- Product Code
- KHO
- PMA / PMN Number
- K971103
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
TOSOH BIOSCIENCE, INC. IS SUBMITTING ON BEHALF OF THE FOREIGN MANUFACTURER, TOSOH CORPORATION, PER EXEMPTION NUMBER E2017013. THIS REPORT IS BEING SUBMITTED DUE TO A RETROSPECTIVE REVIEW CONDUCTED UNDER CAPA-(B)(4). ON 17-JAN-2017, A FIELD SERVICE ENGINEER (FSE) CONDUCTED AN ONSITE FOLLOW-UP TO ADDRESS THE REPORTED EVENT. DURING SERVICING, FSE FOUND THAT THE IMPASSE TEST WAS SLIGHTLY OFF. FSE TRIED REPLACING THE SAMPLE NOZZLE AND RERAN THE SAMPLES BUT OBTAINED THE SAME RESULTS. THEY REPLACED THE SAMPLE NOZZLE WITH AN OLD NOZZLE AND RAN CHECKS ON IMPASSE VALUES. THE VOLTAGE FOR THE IMPASSE SENSOR WAS ADJUSTED AND WAS RAN 5 TIMES WITH SIMILAR RESULTS TO C-B. THE FINAL RESULT WAS NORMAL. THE FSE PERFORMED A SOFTWARE UPGRADE AND THE INSTRUMENT WAS VERIFIED TO BE OPERATING AS EXPECTED. THERE WAS NO FURTHER ACTION BY FSE. THE MOST PROBABLE CAUSE OF THE REPORTED EVENT WAS INCORRECT IMPASSE VOLTAGE.
ON (B)(6) 2017, A CUSTOMER REPORTED ERROR 4017 CLOG DETECTED CLOG B PULSE 15 WITH THEIR AIA-360 ANALYZER. THE CUSTOMER REPORTED THAT THE SAMPLE PROBE IS STILL INTACT BUT THEY ARE GETTING INTERMITTENT POUR OFF INTO TEST CUPS. THEY ARE UNABLE TO RUN PARATHYROID (PTH) PATIENT SAMPLES. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ADDRESS THE REPORTED EVENT, WHICH RESULTED IN DELAYED REPORTING PTH PATIENT RESULTS. THERE WAS NO INDICATION OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DELAY IN REPORTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 179821 | A1A-360 | A1A-360 | KHO | TOSOH CORPORATION | A1A-360 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |