FDA Adverse Event Malfunction Summary report: N

CATALYS SYSTEM

MDR report key: 7336728 · Received March 13, 2018

Report

Report Number
3005675890-2018-00010
Event Type
Malfunction
Date Received
March 13, 2018
Report Date
April 23, 2018
Manufacturer
JOHNSON & JOHNSON SURGICAL VISION, INC
Product Code
OOE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(6). THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT; LASER CORRECTION SYSTEM, MODEL CATALYS-I, THAT HAS A SIMILAR PRODUCT, CATALYS-U WHICH IS DISTRIBUTED IN THE UNITED STATES UNDER 510(K) # K121091. MANUFACTURE DATE - 2014. FIELD SERVICE SPECIALIST VISITED THE ACCOUNT AND CHECKED LASER. HE REPLACED VACUUM REGULATOR DUE TO BALANCED SALT SOLUTION INGRESS. VACUUM CALIBRATION WAS DONE. INVESTIGATION SUBSEQUENTLY IDENTIFIED THAT THE CATALYS SYSTEM MAY HAVE A REMOTE POTENTIAL EVENT WHERE LOSS OF SUCTION DURING LASER FIRING COULD CREATE A HAZARDOUS SITUATION THAT MAY RESULT IN SCORING OF THE POSTERIOR CORNEAL SURFACE. THEREFORE JOHNSON & JOHNSON SURGICAL VISION, INC HAS ISSUED AN ADVISORY NOTICE TO REINFORCE INSTRUCTIONS PROVIDED IN PRODUCT TRAINING AND IN THE OPERATOR¿S MANUAL REGARDING SUCTION LOSS DURING A PROCEDURE. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: A REVIEW OF THE RECORDS RELATED TO THE DEVICE THAT INCLUDED LABELING, MANUALS, TRENDING, AND RISK DOCUMENTATION WAS PERFORMED. THE TREND REVIEW SHOWS THAT THERE IS NOT SIGNIFICANT CHANGE OVER HISTORICAL COMPLAINT LIMITS AND NO RECOGNIZABLE ADVERSE TREND. THE RISKS AND MITIGATIONS ASSOCIATED WITH THE COMPLAINT ISSUE ARE IDENTIFIED IN EXISTING RISK DOCUMENTS AND NO NEW RISKS WERE IDENTIFIED AS PART OF THIS INVESTIGATION. THERE WAS NO PRODUCT DEFICIENCY IDENTIFIED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A CATARACT PROCEDURE AND WHILE THE LASER WAS FIRING THE SYSTEM EXPERIENCED A SUCTION LOSS. THEY REPORTED MULTIPLE ERRORS 02-006 DURING LAST SURGERY. VACUUM SLOW DUE TO BALANCED SALT SOLUTION INGRESS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
178098 CATALYS SYSTEM CATALYS OOE JOHNSON & JOHNSON SURGICAL VISION, INC CATALYS-I

Patients

Seq Age Sex Outcome Treatment
1