CATALYS SYSTEM
Report
- Report Number
- 3006695864-2018-00517
- Event Type
- Malfunction
- Date Received
- March 13, 2018
- Date of Event
- February 12, 2018
- Report Date
- May 2, 2018
- Manufacturer
- JOHNSON & JOHNSON SURGICAL VISION, INC
- Product Code
- OOE
- UDI-DI
- 05050474609792
- PMA / PMN Number
- K121091
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- NURSE
Narratives
DEVICE EVALUATION: A REVIEW OF THE RECORDS RELATED TO THE DEVICE THAT INCLUDED LABELING, MANUALS, TRENDING, AND RISK DOCUMENTATION WAS PERFORMED. THE TREND REVIEW SHOWS THAT THERE IS NOT SIGNIFICANT CHANGE OVER HISTORICAL COMPLAINT LIMITS AND NO RECOGNIZABLE ADVERSE TREND. THE RISKS AND MITIGATION ASSOCIATED WITH THE COMPLAINT ISSUE ARE IDENTIFIED IN EXISTING RISK DOCUMENTS AND NO NEW RISKS WERE IDENTIFIED AS PART OF THIS INVESTIGATION. THERE WAS NO PRODUCT DEFICIENCY IDENTIFIED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.
JOHNSON & JOHNSON SURGICAL VISION, INC. INVESTIGATION SUBSEQUENTLY IDENTIFIED THAT THE CATALYS SYSTEM MAY HAVE A REMOTE POTENTIAL EVENT WHERE LOSS OF SUCTION DURING LASER FIRING COULD CREATE A HAZARDOUS SITUATION THAT MAY RESULT IN SCORING OF THE POSTERIOR CORNEAL SURFACE. THEREFORE JOHNSON & JOHNSON SURGICAL VISION, INC HAS ISSUED AN ADVISORY NOTICE TO REINFORCE INSTRUCTIONS PROVIDED IN PRODUCT TRAINING AND IN THE OPERATOR¿S MANUAL REGARDING SUCTION LOSS DURING A PROCEDURE. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.
IT WAS REPORTED THAT DURING A CATARACT PROCEDURE AND WHILE THE LASER WAS FIRING THE SYSTEM EXPERIENCED A SUCTION LOSS. THE PROCEDURE WAS ABLE TO BE COMPLETED SUCCESSFULLY AND THERE WAS NO SURGICAL INTERVENTION REQUIRED. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 176722 | CATALYS SYSTEM | CATALYS | OOE | JOHNSON & JOHNSON SURGICAL VISION, INC | CATALYS-U | 05050474609792 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |