AIA-600II
Report
- Report Number
- 8031673-2018-01655
- Event Type
- Malfunction
- Date Received
- March 9, 2018
- Date of Event
- February 8, 2017
- Report Date
- March 9, 2018
- Manufacturer
- TOSOH CORPORATION
- Product Code
- KHO
- PMA / PMN Number
- K971103
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
ON (B)(6) 2017 VIA A PHONE CONVERSATION, THE CUSTOMER INFORMED THE TECHNICAL SUPPORT SPECIALIST (TSS) THAT THE BIORAD QC LOT 40330 WAS 245, 621, 1316 PG/ML. ALL QUALITY CONTROL RUNS WERE NOTED TO BE WITHIN QC TARGET RANGE BY TSS. E2 WAS LAST CALIBRATED ON (B)(6) 2017 WITH CUP LOT GX129A1/CAL LOT G632971. TSS DISCUSSED RUNNING A CORRELATION STUDY BETWEEN THE TWO METHODS TO DETERMINE THE RELATIONSHIP BETWEEN THE TWO METHODOLOGIES. THE SECOND METHOD IS UNKNOWN AT THIS TIME. THE PROBABLE CAUSE OF THE EVENT WAS LACK OF OPERATOR UNDERSTANDING OF THE TWO TEST METHODOLOGIES. TOSOH BIOSCIENCE, INC. IS SUBMITTING ON BEHALF OF THE FOREIGN MANUFACTURER, TOSOH CORPORATION, PER EXEMPTION NUMBER E2017013. THIS REPORT IS BEING SUBMITTED DUE TO A RETROSPECTIVE REVIEW CONDUCTED UNDER CAPA(B)(4).
ON (B)(6) 2017 A CUSTOMER REPORTED CONCERNS WITH THEIR E2 RESULTS WITH THEIR AIA-600II. WHEN THE SAMPLES WERE SENT OUT THE RESULTS CAME BACK DIFFERENT. AN EXAMPLE PROVIDED BY THE SITE REVEALED "E2 - 647 PG/ML AND WHEN SENT OUT IT CAME BACK AS 406 PG/ML." THE CUSTOMER REPORTED DISCREPANT PATIENT RESULTS. THE REPRESENTATIVE FROM THE TECHNICAL SUPPORT SPECIALIST (TSS) SPOKE TO THE CUSTOMER TO TROUBLESHOOT THE REPORTED EVENT WHICH RESULTED IN DISCREPANT RESULTS. THERE WAS NO INDICATION OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCREPANT RESULT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 169313 | AIA-600II | AIA-600II | KHO | TOSOH CORPORATION | AIA-600II |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |