FDA Adverse Event Malfunction Summary report: N

ONLY F/MODULAR MONOPOL.ELECTR

MDR report key: 7325381 · Received March 8, 2018

Report

Report Number
9610612-2018-00082
Event Type
Malfunction
Date Received
March 8, 2018
Report Date
March 12, 2018
Manufacturer
AESCULAP AG
Product Code
GEI
PMA / PMN Number
K970541
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER: E2014018. MANUFACTURING SITE EVALUATION: EVALUATION ON-GOING.

Additional Manufacturer Narrative · 0

INVESTIGATION: THE INVESTIGATION WAS CARRIED OUT VISUALLY. THE WORKING END OF THE INSULATION SHAFT IS BURNT DUE TO AN ELECTRIC BREAKDOWN. CONCLUSION AND ROOT CAUSE: BSED ON THE INFORMATION AVALIABLE AS WELL A RESULT OF OUR INVESTIGATION, THE ROOT CAUSE OF THE FAILURE IS MOST PROBABLY RELATED TO AN INSUFFICIENT USAGE. RATIONAL: MOST LIKELY MOISTURE ENTERED THROUGH AN EXISTING HAIR CRACK AND CAUSED AN INSULATION BREAKDOWN. THIS, IN TURN, LED TO AN ELECTRIC ARC, WHICH CAUSED THE DAMAGE OF THE OUTER TUBE. THE SHAFT MUST HAVE BEEN DAMAGED AFTER DELIVERY OF THE INSTRUMENTS. BASED ON THE INSPECTION OF THE QUALITY ASSURANCE, THE SHAFTS ARE INSPECTED FOR CRACKS 100% NO CAPA IS NECESSARY

Description of Event or Problem · 0

(B)(6). PLASTIC SHAFT HAS BURNED AND DAMAGED AT THE TOP DURING LAPAROSCOPIC INTERVENTION AND CAUSED MINOR PATIENT INJURY, PRODUCT IN WARRANTY. COMPONENTS IN USE LISTED AS CONCOMITANT DEVICES ARE: GK370P / SHAFT ONLY F/MODULAR MONOPOL.ELECTR, GK383R / CERAMIC ELECTRODE TIP J-HK F/GK372R, GK373R / INNER TUBE W/HANDLE FOR GK372R.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
166297 ONLY F/MODULAR MONOPOL.ELECTR ENDOSCOPY GEI AESCULAP AG GK370P

Patients

Seq Age Sex Outcome Treatment
1 Other GK370P / SHAFT ONLY F/MODULAR MONOPOL.ELECTR| GK373R / INNER TUBE W/HANDLE FOR GK372R| GK383R / CERAMIC ELECTRODE TIP J-HK F/GK372R