VITROS 5600 INTEGRATED SYSTEM
Report
- Report Number
- 1319681-2018-00025
- Event Type
- Malfunction
- Date Received
- March 7, 2018
- Date of Event
- February 11, 2018
- Report Date
- March 7, 2018
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS
- Product Code
- JJE
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION DETERMINED THAT HIGHER THAN EXPECTED VITROS PHYT RESULTS WERE OBTAINED FROM A VITROS PERFORMANCE VERIFIER QUALITY CONTROL SAMPLE USING A VITROS 5600 INTEGRATED SYSTEM. THE MOST LIKELY ASSIGNABLE CAUSE OF THIS EVENT WAS INSTRUMENT RELATED. THE IMMUNO-WASH FLUID (IWF) TIP WAS BENT, AND THE IWF RESERVOIR COVER WAS NOT IN POSITION. IT IS LIKELY THE IWF TIP HIT THE ¿OUT OF POSITION¿ RESERVOIR COVER, CAUSING THE TIP TO BEND, WHICH IN RETURN DELIVERED AN INSUFFICIENT VOLUME OF IWF. ACCEPTABLE VITROS PHYT PERFORMANCE WAS OBTAINED AFTER REPLACING THE IWF TIP AND FLUID. THE INVESTIGATION DETERMINED THAT THE MOST LIKELY ASSIGNABLE CAUSE OF THIS EVENT WAS INSTRUMENT RELATED.
THE CUSTOMER OBTAINED HIGHER THAN EXPECTED PHYT QUALITY CONTROL RESULTS USING A VITROS 5600 INTEGRATED SYSTEM (UNITS ARE ¿G/ML): TDM PV L3 LOT T5896 VITROS PHYT RESULTS OF >0.40 (X4) ¿G/ML VERSUS EXPECTED VALUE OF 28.1 ¿G/ML. BIASED RESULTS OF THE DIRECTION AND MAGNITUDE OBSERVED COULD LEAD TO INAPPROPRIATE PHYSICIAN ACTION. THERE WAS NO REPORT THAT PATIENT SAMPLES WERE AFFECTED AND NO ALLEGATION OF PATIENT HARM AS A CAUSE OF THIS EVENT. HOWEVER, THE INVESTIGATION CANNOT CONCLUDE THAT PATIENT SAMPLES WOULD NOT BE AFFECTED IF THE EVENT WERE TO RECUR UNDETECTED. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 165573 | VITROS 5600 INTEGRATED SYSTEM | CHEMISTRY ANALYZER | JJE | ORTHO-CLINICAL DIAGNOSTICS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |