HEARTSINE SAMARITAN 450P AND PAD-PAK
Report
- Report Number
- 3004123209-2018-00209
- Event Type
- Malfunction
- Date Received
- March 7, 2018
- Date of Event
- February 26, 2018
- Report Date
- May 29, 2018
- Manufacturer
- HEARTSINE TECHNOLOGIES LTD
- Product Code
- MKJ
- PMA / PMN Number
- P16008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). H3 OTHER TEXT : DEVICE NOT RETURNED YET.
THE DEVICE HISTORY RECORDS FOR THE SAM 450P DEVICE AND PAD-PAK WERE REVIEWED AND THIS CONFIRMED THAT ALL MANUFACTURING AND QUALITY CHECKS AND TEST HAD BEEN SUCCESSFULLY COMPLETED. NO REWORK WAS CONDUCTED. THE SAM 450P PASSED ¿OUT QAT FROM HEARTSINE TECHNOLOGIES ON THE 14TH MAY 2015. THE DEVICE PERFORMED TO SPECIFICATION THROUGHOUT TESTING AT HEARTSINE. ON THE (B)(6) 2018, INFORMATION FROM THE TECHNICAL LOG INDICATES THE SHOCK KEY HAD EITHER BEEN ¿STUCK¿ OR HELD DOWN BY AN EXTERNAL FORCE, AS THE DEVICE RECORDS A FAILED WEEKLY AUTO SELF-TEST, AS PER THE REPORTED FAULT. THIS RESULTED IN THE DEVICE POWERING OFF WITH A FLASHING RED STATUS LED AND A ¿WARNING DEVICE SERVICE REQUIRED¿ MESSAGE ALONG WITH A FAILURE CHIRP. THE DEVICE WAS SUCCESSFULLY POWER CYCLED PASSING A SELF-TEST A DAY LATER. NO FAULT COULD BE MEASURED OR OBSERVED WITH THE SHOCK KEY LINE. THE DEVICE WAS TEMPERATURE CYCLED BETWEEN 0-50°C FOR 48 HOURS WHILE PERFORMING A SELF-TEST EVERY 20 MINUTES, THIS EQUATES TO APPROXIMATELY 33 MONTHS OF NORMAL OPERATION WITHOUT FAULT. THE FAULT WAS ONLY REPLICATED BY HOLDING IN THE SHOCK KEY DURING A SELF-TEST AT THE START OF A MANUAL POWER CYCLE. IT IS THE POLICY OF HEARTSINE NOT TO REFURBISH DEVICES WHICH HAVE BEEN RETURNED FROM THE FIELD, AFTER INVESTIGATION, THEREFORE THIS DEVICE SHALL BE SCRAPPED AND REPLACED WITH ANOTHER DEVICE OF THE SAME MODEL. DEVICE NOT RETURNED YET.
SAM450P UNIT FAILS IT'S WEEKLY SELF TEST WITH PAD PAK EXPIRY 8-2021. NO PATIENT INVOLVED.
SAM450P UNIT FAILS IT'S WEEKLY SELF TEST WITH PAD PAK EXPIRY 8-2021. NO PATIENT INVOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 165474 | HEARTSINE SAMARITAN 450P AND PAD-PAK | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES LTD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |