FDA Adverse Event
Malfunction
Summary report: N
COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE
MDR report key: 731914
·
Received June 23, 2006
Report
- Report Number
- 2023826-2006-00860
- Event Type
- Malfunction
- Date Received
- June 23, 2006
- Date of Event
- May 30, 2006
- Report Date
- May 31, 2006
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- HQL
- PMA / PMN Number
- p990013
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
H6: EVALUATION RESULTS - VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND A PIECE OF THE LENS OPTIC AND ONE HAPTIC TORN OFF AND MISSING. THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. CONCLUSIONS - (NO CONCLUSION CAN BE DRAWN): BASED ON THE COMPLAINT HISTORY AND THE EVALUATION OF THE RETURNED PRODUCT, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. IT SHOULD BE NOTED THAT THE INJECTOR AND CARTRIDGE WERE NOT RETURNED FOR EVALUATION. PMA# P990013.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE SURGEON INSERTED A CQ2015 COLLAMER THREE-PIECE LENS AND ONE HAPTIC TORE. THE LENS WAS REMOVED WITH NO PATIENT INJURY, NO ENLARGED INCISION OR SUTURES. ANOTHER SAME TYPE LENS WAS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE | INTRAOCULAR LENS | HQL | STAAR SURGICAL CO. | CQ2015 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | VISCOCLASTIC TYPE STAARVISC II - LOT NUMBER UNK| INJECTOR MODEL MSI-TM - LOT NUMBER UNK| CARTRIDGE MODEL CQ CARTRIDGE-FP- LOT NUMBER UNK |