CLARIVEIN IC INFUSION CATHETER
Report
- Report Number
- 3005831739-2018-00013
- Event Type
- Malfunction
- Date Received
- March 6, 2018
- Date of Event
- October 17, 2017
- Report Date
- March 6, 2018
- Manufacturer
- VASCULAR INSIGHTS LLC
- Product Code
- KRA
- UDI-DI
- 10854339004027
- PMA / PMN Number
- K071468
- Removal / Correction Number
- 3005831739-11-09-17-123R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS FAILURE MODE, DIFFICULTY OF INSERTION, WAS FULLY INVESTIGATED AND WAS ISOLATED TO ONE LOT OF PRODUCT AS IDENTIFIED IN THIS REPORT. ROOT CAUSE ANALYSIS WAS PERFORMED AND CORRECTIVE ACTIONS IDENTIFIED AND IMPLEMENTED. SINCE THE IMPLEMENTATION OF CORRECTIVE ACTION THERE HAVE BEEN NO FURTHER EVENTS OF THIS NATURE REPORTED.
THE DOCTOR PREPARED THE PATIENT FOR THE PROCEDURE BY INSERTING THE INTRODUCER SHEATH. PHYSICIAN REMOVED THE CLARIVEIN CATHETER PART NUMBER (B)(4) FROM THE PACKAGING AND PROCEEDED TO INSERT THE CATHETER ASSEMBLY INTO THE INTRODUCER SHEATH. THE PHYSICIAN ENCOUNTERED DIFFICULTY ADVANCING THE CATHETER INTO THE INTRODUCER SHEATH. PHYSICIAN DISCARDED THE DEVICE AND OPENED A NEW ONE TO COMPLETE THE PROCEDURE WITH NO OTHER ISSUES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 160487 | CLARIVEIN IC INFUSION CATHETER | CLARIVEIN IC INFUSION CATHETER | KRA | VASCULAR INSIGHTS LLC | 65-018-E4S | BM651403017I | 10854339004027 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |