FDA Adverse Event Malfunction Summary report: N

OFFSET CUP REAMER HANDLE

MDR report key: 7316023 · Received March 6, 2018

Report

Report Number
3005985723-2018-00136
Event Type
Malfunction
Date Received
March 6, 2018
Date of Event
February 8, 2018
Report Date
May 3, 2018
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
PMA / PMN Number
K170593
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

FOLLOW-UP #1 AND FINAL REPORT SUBMITTED. REPORTED EVENT: THE EVENT REPORTED THAT 4 OFFSET REAMER HANDLES AND 4 TRIDENT INLINE-OFFSET IMPACTORS HAD BIOFILM BUILD UP. THE PROCEDURE WAS DELAYED 5 MINUTES. DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION SHOWED NO FOREIGN MATTER (BIOFILM) OR DEBRIS. THE YOKES OF THE PROXIMAL U-JOINT WAS FOUND TO BE WORN AT THE EDGES. FUNCTIONAL INSPECTION SHOWED THAT THE DEVICE ROTATED FREELY AND SMOOTHLY. DEVICE HISTORY REVIEW: REVIEW OF DEVICE HISTORY RECORDS SHOW THAT 22 DEVICES WERE ACCEPTED INTO FINAL STOCK AS "USE AS IS" PER QT 17-07-0031 ON 07/13/2017. THE NONCONFORMANCE WAS UNRELATED TO THE FAILURE IN THIS INVESTIGATION. COMPLAINT HISTORY REVIEW: REVIEW OF COMPLAINTS FOR P/N 212760, L/N 3761448 SHOW NO OTHER COMPLAINT RELATED TO THE FAILURE IN THIS COMPLAINT. CONCLUSION: THE FAILURE WAS NOT CONFIRMED. BIOFILM WAS NOT SEEN WITHIN THE DEVICE. CORRECTIVE ACTION / PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD.

Description of Event or Problem · 0

BIOFILM BUILD UP DISCOVERED INSIDE THE OFFSET REAMERS AND IN LINE OFFSET IMPACTORS. THA PROCEDURE DELAYED MORE THAN 5 MINUTES.

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

BIOFILM BUILD UP DISCOVERED INSIDE THE OFFSET REAMERS AND IN LINE OFFSET IMPACTORS. THA PROCEDURE DELAYED MORE THAN 5 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
160567 OFFSET CUP REAMER HANDLE STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. 3761448

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization