OFFSET CUP REAMER HANDLE
Report
- Report Number
- 3005985723-2018-00136
- Event Type
- Malfunction
- Date Received
- March 6, 2018
- Date of Event
- February 8, 2018
- Report Date
- May 3, 2018
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- PMA / PMN Number
- K170593
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP #1 AND FINAL REPORT SUBMITTED. REPORTED EVENT: THE EVENT REPORTED THAT 4 OFFSET REAMER HANDLES AND 4 TRIDENT INLINE-OFFSET IMPACTORS HAD BIOFILM BUILD UP. THE PROCEDURE WAS DELAYED 5 MINUTES. DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION SHOWED NO FOREIGN MATTER (BIOFILM) OR DEBRIS. THE YOKES OF THE PROXIMAL U-JOINT WAS FOUND TO BE WORN AT THE EDGES. FUNCTIONAL INSPECTION SHOWED THAT THE DEVICE ROTATED FREELY AND SMOOTHLY. DEVICE HISTORY REVIEW: REVIEW OF DEVICE HISTORY RECORDS SHOW THAT 22 DEVICES WERE ACCEPTED INTO FINAL STOCK AS "USE AS IS" PER QT 17-07-0031 ON 07/13/2017. THE NONCONFORMANCE WAS UNRELATED TO THE FAILURE IN THIS INVESTIGATION. COMPLAINT HISTORY REVIEW: REVIEW OF COMPLAINTS FOR P/N 212760, L/N 3761448 SHOW NO OTHER COMPLAINT RELATED TO THE FAILURE IN THIS COMPLAINT. CONCLUSION: THE FAILURE WAS NOT CONFIRMED. BIOFILM WAS NOT SEEN WITHIN THE DEVICE. CORRECTIVE ACTION / PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD.
BIOFILM BUILD UP DISCOVERED INSIDE THE OFFSET REAMERS AND IN LINE OFFSET IMPACTORS. THA PROCEDURE DELAYED MORE THAN 5 MINUTES.
AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
BIOFILM BUILD UP DISCOVERED INSIDE THE OFFSET REAMERS AND IN LINE OFFSET IMPACTORS. THA PROCEDURE DELAYED MORE THAN 5 MINUTES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 160567 | OFFSET CUP REAMER HANDLE | STEREOTAXIC DEVICE, ROBOTICS | OLO | MAKO SURGICAL CORP. | 3761448 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |