FDA Adverse Event Malfunction Summary report: N

RELION® INSULIN SYRINGE

MDR report key: 7311196 · Received March 2, 2018

Report

Report Number
1920898-2018-00142
Event Type
Malfunction
Date Received
March 2, 2018
Date of Event
January 1, 2018
Report Date
March 5, 2018
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
UDI-DI
00681131311748
PMA / PMN Number
K170386
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: SEVERITY: S_1__; OCCURRENCE: A COMPLAINT HISTORY CHECK WAS PERFORMED AND THIS IS THE 6TH RELATED COMPLAINT REPORTED FOR THE DEFECT/CONDITION ON LOT NUMBER 7023647. INVESTIGATION SUMMARY: CUSTOMER RETURNED (2) 1CC, 8MM, 31G RELION SYRINGES IN AN OPEN POLY BAG FROM LOT # 7023647. CUSTOMER STATES THAT WHEN REMOVING THE SYRINGE THE NEEDLE BREAKS OFF OF THE HUB AND COMES OFF INSIDE OF THE SYRINGE. BOTH RETURNED SYRINGES WERE EXAMINED AND NO DEFECTS WERE OBSERVED ON EITHER OF THE SAMPLES. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH # 7023647. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE SEVEN (7) NOTIFICATIONS [(B)(4)] NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND NO CAPA IS REQUIRED AT THIS TIME.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION AND/OR DEVICE HISTORY REVIEW, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE NEEDLE DETACHED FROM THE HUB OF A RELION® INSULIN SYRINGE. THERE WAS NO REPORT OF EXPOSURE, INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
154799 RELION® INSULIN SYRINGE INSULIN SYRINGE FMF BD MEDICAL - DIABETES CARE 7023647 00681131311748

Patients

Seq Age Sex Outcome Treatment
1 Other