IPC® HANDPIECE - INDIGO¿ DRILL
Report
- Report Number
- 1045254-2018-00080
- Event Type
- Malfunction
- Date Received
- March 2, 2018
- Date of Event
- February 7, 2018
- Report Date
- April 2, 2018
- Manufacturer
- MEDTRONIC XOMED, INC.
- Product Code
- HBE
- UDI-DI
- 00613994933713
- PMA / PMN Number
- K081475
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
CONCOMITANT DEVICE: INDIGO ATTACHMENT 1845020 , S/N: (B)(4) , LOT NO: 207410087, MANUFACTURING DATE 2013-09-11. 510K: K081475. THE PRODUCTS HAVE BEEN RETURNED. ANALYSIS OF THE DEVICES IS PENDING. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
ANALYSIS OF THE DEVICES HAS BEEN COMPLETED. DRILL 1845000 INDIGO HIGH SPEED SERIAL NUMBER (B)(4): COMPLAINT HAS NOT BEEN CONFIRMED. TEMPERATURE TEST AT ONE MINUTE: T1 ¿ 79; T2 ¿ 80. THE DEVICE HAS BEEN SUCCESSFULLY TESTED. CONCOMITANT: INDIGO ATTACHMENT ANGLED 1845020 , S/N: (B)(4) , LOT NO: 207410087: CUSTOMER COMPLAINT CANNOT BE CONFIRMED, TEMPERATURE AFTER 5 MINUTES: 103 F. THIS IS WITHIN OUR SPECIFICATIONS. THE 1845020 DEVICE HAS BEEN SUCCESSFULLY TESTED. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE CUSTOMER REPORTED THAT THE DEVICE GETS HOT REALLY QUICKLY DURING USE (IN A FEW SECONDS) AND REGARDING THE ATTACHMENT THAT THEY WERE NOT SURE WHETHER IT'S BECAUSE DRILL GETS HOT, OR BECAUSE THERE'S SOMETHING WRONG WITH THE BEARINGS. THERE WAS NO USER/PATIENT IMPACT OR INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 154136 | IPC® HANDPIECE - INDIGO¿ DRILL | DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) | HBE | MEDTRONIC XOMED, INC. | 1845000 | NI | 00613994933713 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |