FDA Adverse Event Injury Summary report: N

AMISTEM H HA COATED STD STEM SIZE6

MDR report key: 7306125 · Received March 1, 2018

Report

Report Number
3005180920-2018-00102
Event Type
Injury
Date Received
March 1, 2018
Date of Event
January 30, 2018
Report Date
March 1, 2018
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030804120
PMA / PMN Number
K093944
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS ON 16 FEBRUARY 2018: HIP REVISION SURGERY OCCURRED 5,5 YEARS AFTER PRIMARY CEMENTLESS TOTAL HIP ARTHROPLASTY. RADIOGRAPHIC IMAGE PROVIDED SHOWS THE PRESENCE OF A MOBILIZED STEM. ASEPTIC LOOSENING IS A POSSIBLE, LITERATURE DESCRIBED ADVERSE EVENT AFTER CEMENTLESS THA AND CAUSES ARE OFTEN UNKNOWN. IN THIS CASE, REASON OF FAILURE CANNOT BE DETERMINED. BATCH REVIEW PERFORMED ON 26 FEBRUARY 2018 LOT 120924: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 22 JUNE 2012. EXPIRATION DATE: 2017-05-31 NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

THE PATIENT CAME IN COMPLAINING OF INSTABILITY 5 YEARS AND 5 MONTHS AFTER PRIMARY. THE CAUSE IS UNKNOWN. THE STEM CAME OUT VERY EASILY, TOOK NO TIME WITH THE MEDACTA STEM EXTRACTOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
151433 AMISTEM H HA COATED STD STEM SIZE6 CEMENTLESS HIP STEM LZO MEDACTA INTERNATIONAL SA 120924 07630030804120

Patients

Seq Age Sex Outcome Treatment
1 49 YR Required Intervention