METRIX AIR
Report
- Report Number
- 1000113657-2018-00246
- Event Type
- Malfunction
- Date Received
- March 1, 2018
- Date of Event
- February 5, 2018
- Report Date
- March 1, 2018
- Manufacturer
- TRIVIDIA HEALTH, INC.
- Product Code
- NBW
- UDI-DI
- 00021292007447
- PMA / PMN Number
- K150052
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4). RETURNED METER EVALUATED WITH NO DEFECT FOUND. TEST STRIPS NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-58-USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST. (B)(4). NOTE: MANUFACTURER CONTACTED CUSTOMER ON 2/20/2018 IN A FOLLOW-UP CALL IN ORDER TO ENSURE THAT THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN; CUSTOMER STATED THAT THE REPLACEMENT PRODUCT IS WORKING TO THEIR SATISFACTION AND THEY HAVE NOT HAD ANY MEDICAL INTERVENTION SINCE THE LAST CALL.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 220, 324, 214 AND 206 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 116 - 138 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. CUSTOMER DECLINED TO PERFORM A BACK TO BACK BLOOD TEST DURING THE CALL ON (B)(6) 2018. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 09/15/2018 AND OPEN VIAL DATE IS (B)(6) 2018. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1 :214MG/DL DATE:(B)(6) 2018 TIME:10:51PM FASTING, RESULT 2 :220MG/DL DATE:(B)(6) 2018 TIME:11:12PM FASTING, RESULT 3 :145MG/DL DATE:(B)(6) 2018 TIME:06:07PM FASTING, RESULT 4 :193MG/DL DATE:(B)(6) 2018 TIME:10:09PM FASTING, RESULT 5 :206MG/DL DATE:(B)(6) 2018 TIME:07:04PM FASTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 149901 | METRIX AIR | BLOOD GLUCOSE SYSTEM | NBW | TRIVIDIA HEALTH, INC. | METRIX AIR | MU2389 | 00021292007447 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |