FDA Adverse Event Malfunction Summary report: N

PFC* MOD PLUS TORQUE DRIVER

MDR report key: 7304309 · Received February 28, 2018

Report

Report Number
1818910-2018-54313
Event Type
Malfunction
Date Received
February 28, 2018
Report Date
January 29, 2018
Manufacturer
DEPUY ORTHOPAEDICS INC 1818910
Product Code
HWR
UDI-DI
10603295215868
PMA / PMN Number
NI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL NARRATIVE: PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: EXAMINATION OF THE RETURNED INSTRUMENT CONFIRMED THE REPORTED EVENT. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT: NULL. DEVICE HISTORY BATCH: NULL. DEVICE HISTORY REVIEW: NULL. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

(B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT #: (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

THE FOLLOWING INSTRUMENTS DIDN¿T FUNCTION CORRECTLY DURING KNEE REVISION SURGERY. THE 860284. LOT UNKNOWN -PFC MOD PLUS TORQUE DRIVER DIDN¿T APPEAR TO CLICK AT MAXIMUM TORQUE CAUSING THE WOBBLE BITS TO FRACTURE. THE 961673. LOT SO2008134. -ADAPTER TORQUE WRENCH. THE BLACK PLASTIC ADAPTER THAT POSITIONS OVER THE FEMORAL ADAPTER DETACHED FROM THE INSTRUMENT DURING TIGHTENING. THE 217863136. LOT TBJGF. - MODIFIED HUDSON ADAPTER GOT STUCK ON ONE OF THE MBT REAMERS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
148179 PFC* MOD PLUS TORQUE DRIVER SCREWDRIVERS HWR DEPUY ORTHOPAEDICS INC 1818910 NI 10603295215868

Patients

Seq Age Sex Outcome Treatment
1