PFC* MOD PLUS TORQUE DRIVER
Report
- Report Number
- 1818910-2018-54313
- Event Type
- Malfunction
- Date Received
- February 28, 2018
- Report Date
- January 29, 2018
- Manufacturer
- DEPUY ORTHOPAEDICS INC 1818910
- Product Code
- HWR
- UDI-DI
- 10603295215868
- PMA / PMN Number
- NI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
ADDITIONAL NARRATIVE: PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: EXAMINATION OF THE RETURNED INSTRUMENT CONFIRMED THE REPORTED EVENT. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT: NULL. DEVICE HISTORY BATCH: NULL. DEVICE HISTORY REVIEW: NULL. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
(B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PRODUCT COMPLAINT #: (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THE FOLLOWING INSTRUMENTS DIDN¿T FUNCTION CORRECTLY DURING KNEE REVISION SURGERY. THE 860284. LOT UNKNOWN -PFC MOD PLUS TORQUE DRIVER DIDN¿T APPEAR TO CLICK AT MAXIMUM TORQUE CAUSING THE WOBBLE BITS TO FRACTURE. THE 961673. LOT SO2008134. -ADAPTER TORQUE WRENCH. THE BLACK PLASTIC ADAPTER THAT POSITIONS OVER THE FEMORAL ADAPTER DETACHED FROM THE INSTRUMENT DURING TIGHTENING. THE 217863136. LOT TBJGF. - MODIFIED HUDSON ADAPTER GOT STUCK ON ONE OF THE MBT REAMERS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 148179 | PFC* MOD PLUS TORQUE DRIVER | SCREWDRIVERS | HWR | DEPUY ORTHOPAEDICS INC 1818910 | NI | 10603295215868 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |