FDA Adverse Event Malfunction Summary report: N

STEALTHSTATION S7 SYSTEM

MDR report key: 7297590 · Received February 26, 2018

Doing adverse event research by hand?

The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.

Free to try · every answer cites its records

Report

Report Number
1723170-2018-00861
Event Type
Malfunction
Date Received
February 26, 2018
Date of Event
January 30, 2018
Report Date
May 16, 2018
Manufacturer
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
Product Code
HAW
UDI-DI
00613994450944
PMA / PMN Number
K050438
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: A MEDTRONIC REPRESENTATIVE WENT TO THE SITE TO TEST THE EQUIPMENT. THE HARDWARE, SOFTWARE, AND INSTRUMENTS PASSED THE SYSTEM CHECKOUT. THE SYSTEM WAS FOUND TO BE FULLY FUNCTIONAL. THE ISSUE COULD NOT BE REPRODUCED.

Additional Manufacturer Narrative · 1

NO PARTS HAVE BEEN RECEIVED BY THE MANUFACTURER FOR EVALUATION. PART NOT RETURNED FOR ANALYSIS

Additional Manufacturer Narrative · 1

THE SOFTWARE INVESTIGATION FOUND THAT THE REPORTED EVENT WAS UNRELATED TO A SOFTWARE ISSUE. THE SOFTWARE FUNCTIONED AS DESIGNED.

Description of Event or Problem · 1

A MEDTRONIC REPRESENTATIVE REPORTED THAT WHILE IN A CRANIAL RESECTION PROCEDURE, THE DYNAMIC REFERENCE FRAME (DRF) PATIENT TRACKER WAS FLICKERING IN AND OUT OF TRACKING STATUS. IT WAS REPORTED THAT THE DRF WAS PLUGGED INTO DIFFERENT PORTS ON THE AXIEM BOX, BUT THE ISSUE OCCURRED EVEN WITH THE EMITTER TRACKING VIEW ADJUSTED. THERE WERE NO ISSUES WITH FLICKERING TRACKING WITH OTHER INSTRUMENTS. TWO DIFFERENT DRF PATIENT TRACKERS WERE TESTED AND BOTH WERE FLICKERING. IT WAS POSSIBLE TO CONTINUE WITH NAVIGATION INSPITE OF THIS. THERE WAS A REPORTED DELAY TO THE PROCEDURE OF 15 MINUTES DUE TO THIS ISSUE AND NO IMPACT ON THE PATIENT OUTCOME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
139358 STEALTHSTATION S7 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW MEDTRONIC NAVIGATION, INC. (LOUISVILLE) S7 00613994450944

Patients

Seq Age Sex Outcome Treatment
1 11 YR