INNOVA 2000
Report
- Report Number
- 9611343-2018-00001
- Event Type
- Injury
- Date Received
- February 26, 2018
- Date of Event
- January 30, 2018
- Report Date
- March 26, 2018
- Manufacturer
- GE MEDICAL SYSTEMS SCS
- Product Code
- OWB
- PMA / PMN Number
- K022322
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- SERVICE PERSONNEL
Narratives
GE HEALTHCARE INVESTIGATION ABOUT THIS EVENT HAS BEEN COMPLETED. ON JANUARY 30, 2018 IT WAS REPORTED THAT DURING MAINTENANCE ACTIVITY ON AN INNOVA 2000 VASCULAR SYSTEM, A GE HEALTHCARE FIELD SERVICE ENGINEER HAD TO REMOVE TABLE COVERS. WHEN REASSEMBLING THE COVERS, HE COULD NOT HOLD THE MOBILE COVER PROPERLY IN THE CONSTRAINED ENVIRONMENT, AND THE COVER SLIPPED FROM HIS HAND. THE FALLING COVER CAME IN CONTACT WITH THE SOFT SKIN OF HIS LEFT HAND PALM BETWEEN THUMB AND INDEX FINGER AND HURT HIS PALM. HE RECEIVED ANTIBIOTICS AND TETANUS SHOT. STITCHES WERE REQUIRED AS WELL AS A RIGID SPLINT FOR THE INJURY HE GOT. GE HEALTHCARE FIELD SERVICE ENGINEER CONFIRMED THAT HE WAS WEARING PROTECTION GLOVES DURING THIS INTERVENTION. IT WAS ALSO CONFIRMED THAT THERE WAS NO SHARP CUTTING EDGES, BURRS OR DAMAGES WITH THE TABLE COVERS. ADEQUATE INSTRUCTIONS ARE PROVIDED IN INSTALLATION MANUAL REGARDING PROPER REMOVAL AND ASSEMBLY OF TABLE COVERS. NEITHER DESIGN, ASSEMBLY NOR MANUFACTURE DEFECT, NOR FAILURE OR MALFUNCTION HAS BEEN IDENTIFIED WITH THE TABLE COVER. THIS IS AN ISOLATED CASE. THIS EVENT IS DUE TO AN IMPROPER HOLDING OF THE COVER BY GE HEALTHCARE FIELD SERVICE ENGINEER. GE HEALTHCARE FIELD SERVICE ENGINEER WAS CONTACTED AND FOUND TO BE AWARE OF TABLE MOBILE COVER FIXING PROCEDURE AND ABOUT THE APPROPRIATE PPE (PERSONNEL PROTECTIVE EQUIPMENT) USAGE. NO FURTHER ACTION IS REQUIRED.
GE HEALTHCARE'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED. DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.
IT HAS BEEN REPORTED THAT DURING TROUBLESHOOTING AN ISSUE ON THE VASCULAR TABLE, GE HEALTHCARE FIELD SERVICE ENGINEER SUSTAINED A LACERATION TO HIS THUMB (OF HIS DOMINANT HAND) WHILE HANDLING A TABLE COVER. GE HEALTHCARE FIELD SERVICE ENGINEER WAS CONDUCTING A REPAIR AND REMOVING THE COVERS WHEN ONE OF THE COVERS FELL AND CUT HIS LEFT THUMB. GE HEALTHCARE FIELD SERVICE ENGINEER RECEIVED ANTIBIOTICS AND TETANUS SHOT. STITCHES WERE REQUIRED AS WELL AS A RIGID SPLINT FOR THE INJURY HE GOT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 139683 | INNOVA 2000 | INTERVENTIONAL FLUOROSCOPIC X-RAY | OWB | GE MEDICAL SYSTEMS SCS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR | Required Intervention |