FDA Adverse Event Malfunction Summary report: N

LMA PROSEAL, REU, SIZE 3 (150030)

MDR report key: 7295691 · Received February 26, 2018

Report

Report Number
9681900-2018-00016
Event Type
Malfunction
Date Received
February 26, 2018
Date of Event
January 19, 2018
Report Date
February 2, 2018
Manufacturer
TELEFLEX MEDICAL
Product Code
CAE
PMA / PMN Number
K780900
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

(B)(4). THE SAMPLE WAS RETURNED FOR EVALUATION. THE DEVICE COULD BE INFLATED NORMALLY WITH THE CHECK VALVE. THE CUFF COULD NOT BE INFLATED WHEN THE RED PLUG WAS OPENED. A DEVICE HISTORY RECORD REVIEW WAS PERFORMED AND THERE WERE NO ISSUES FOUND THAT COULD RELATE TO THE REPORTED COMPLAINT. THE OPENING OF THE RED PLUG WAS BLOCKED BY THE VALVE HENCE THE CUFF COULD NOT BE DEFLATED EVEN WHEN THE RED PLUG WAS OPENED. THE MOST PROBABLE ROOT CAUSE IS THAT EXCESSIVE FORCE WAS APPLIED TO THE VALVE INSIDE THE BLUE INFLATION BALLOON WHEN INFLATING OR DEFLATING THE DEVICE. THE VALVE HAD SUNK FURTHER INTO THE TUBING WHICH RESULTED IN THE RED PLUG OPENING BEING BLOCKED BY THE VALVE. NONETHELESS, THE DEVICE IS STILL FUNCTIONAL AND CAN STILL BE INFLATED AND DEFLATED USING THE CHECK VALVE.

Description of Event or Problem · 0

CUSTOMER ALLEGES THE CUFF DOES NOT INFLATE. ALLEGED DEFECT REPORTED AS "FOUND AT THE PRE-USE INSPECTION". CUSTOMER REPORTS THE DEVICE WAS "NOT CLINICALLY USED, MEANING NO PATIENT AFFECTED FROM THIS COMPLAINT."

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE INVOLVED IN THIS COMPLAINT HAS NOT BEEN RETURNED AT THE TIME OF THIS REPORT. THE INVESTIGATION INTO THIS COMPLAINT IS STILL IN PROGRESS.

Description of Event or Problem · 1

CUSTOMER ALLEGES THE CUFF DOES NOT INFLATE. ALLEGED DEFECT REPORTED AS "FOUND AT THE PRE-USE INSPECTION". CUSTOMER REPORTS THE DEVICE WAS "NOT CLINICALLY USED, MEANING NO PATIENT AFFECTED FROM THIS COMPLAINT."