FDA Adverse Event Malfunction Summary report: N

ULTRACISION SURG DEV, CURVED SHEARS, ENDO

MDR report key: 729545 · Received June 19, 2006

Report

Report Number
1527736-2006-02771
Event Type
Malfunction
Date Received
June 19, 2006
Date of Event
November 8, 2005
Report Date
February 15, 2006
Manufacturer
ETHICON ENDO SURGERY, INC.(CINCINNATI)
Product Code
LFL
PMA / PMN Number
k002981
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BASED ON ANALYSIS RESULTS, THIS COMPLAINT IS NOW DETERMINED TO BE A MDR MALFUNCTION. THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED. THE DEVICE WAS TESTED WITH A GENERATOR AND A SOLID TONE WAS RECEIVED. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FROM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. THEREFORE, THE INSTRUCTIONAL INSERT STATES: "SCRATCHES ON THE BLADE MAY LEAD TO PREMATURE BLADE FAILURE" AND "AVOID ACCIDENTAL CONTACT WITH OTHER INSTRUMENTS DURING USE." THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MFG PROCESS. 510(K) NUMBER IS K002981.

Description of Event or Problem · 1

PRIOR TO AN UNK PROCEDURE, ERROR CODE 5: INSTRUMENT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PT. ONE DEVICE IS RETURNING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION SURG DEV, CURVED SHEARS, ENDO GEI LFL ETHICON ENDO SURGERY, INC.(CINCINNATI) NA A4AM39

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN 1.GENERATOR, 2.HAND PIECE.