ADVIA 1800 ANALYZER
Report
- Report Number
- 2432235-2018-00087
- Event Type
- Malfunction
- Date Received
- February 23, 2018
- Date of Event
- February 1, 2018
- Report Date
- February 23, 2018
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- JJE
- UDI-DI
- 00630414553177
- PMA / PMN Number
- K990346
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A SIEMENS CUSTOMER SERVICE ENGINEER (CSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE CSE DECONTAMINATED THE INSTRUMENT AND REBUILT THE PUMPS. THE CSE FOUND THE REACTION TRAY WASH UNIT DRAIN VALVE 1 (CDEV1) ON THE REACTION TRAY WASH UNIT (WUD) NOT OPENING AND REPLACED THE VALVE; CLEANED AND CHECKED FILTRATION OF MIX 1 WASH PORT, CHECKED WUD ASPIRATORS, REPLACED THE LARGE WATER PUMP (LWP) FILTER, AND ADJUSTED THE ALIGNMENT OF THE REAGENT PIPETTING PROBE 1 (RPP1) WASH PORT. THE CSE CLEANED AND VERIFIED FULL DRAIN OF RPP1 WASH PORT, INSTALLED THE EXTERNAL DEGASSER AND RAN QC AS WELL AS MULTIPLE SAMPLES. ALL PASSED AND NO OUTLIERS WERE SEEN. THE CAUSE OF THE DISCORDANT, FALSELY ELEVATED CARBON DIOXIDE CONCENTRATED (CO2_C) ON THE TWO PATIENT SAMPLES IS UNKNOWN. NO FURTHER ISSUES WITH CO2_C AFTER SERVICE WAS DISPATCHED. THE DEVICE IS PERFORMING WITHIN MANUFACTURING SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
DISCORDANT, FALSELY ELEVATED CARBON DIOXIDE CONCENTRATED (CO2_C) RESULTS WERE OBTAINED ON 2 SAMPLES ON AN ADVIA 1800 INSTRUMENT. THE DISCORDANT RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLES WERE REPEATED ON THE SAME INSTRUMENT AND THEY RESULTED LOWER. THE CORRECTED RESULTS WERE REPORTED TO THE PHYSICIAN(S). THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED CARBON DIOXIDE CONCENTRATED RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 136905 | ADVIA 1800 ANALYZER | ADVIA CHEMISTRY | JJE | SIEMENS HEALTHCARE DIAGNOSTICS INC. | ADVIA 1800 | 00630414553177 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |