ANIMAS VIBE
Report
- Report Number
- 2531779-2018-04093
- Event Type
- Malfunction
- Date Received
- February 22, 2018
- Report Date
- February 13, 2018
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- OYC
- PMA / PMN Number
- P130007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER CAREGIVERS
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1: DATE OF SUBMISSION: 06-APR-2018. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 28-MAR-2018 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE BLACK BOX AND ALARM HISTORY SHOW NO EVIDENCE OF ¿CS120-0830042¿ ALARM. RETURNED BATTERY CAP IS UNDAMAGED AND ABLE TO FIT THE EZ-PRIME¿ STEPS WERE PERFORMED CORRECTLY, NO ALARMS OCCURRED DURING THE INVESTIGATION, UNABLE TO DUPLICATE ¿CS120-0830042¿ COMPLAINT. PUMP¿S COVER WAS REMOVED, NO MOISTURE CORROSION OR INTERMITTENT CONDITION WAS FOUND TO THE POWER PRINTED CIRCUIT BOARD. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2018, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CALL SERVICE ALARM ISSUE) ISSUE. IT WAS ALLEGED A CALL SERVICE 120 ALARM WAS EMITTED. THERE WAS NO INDICATION THAT THIS ISSUE LEAD TO AN ADVERSE EVENT. THIS IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 131581 | ANIMAS VIBE | INSULIN INFUSION PUMP | OYC | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR |