FDA Adverse Event Malfunction Summary report: N

IMMUNOSCAN CCPLUS

MDR report key: 7288168 · Received February 22, 2018

Report

Report Number
9616003-2018-00003
Event Type
Malfunction
Date Received
February 22, 2018
Date of Event
February 26, 2016
Report Date
February 21, 2018
Manufacturer
EURO DIAGNOSTICA AB
Product Code
NHX
UDI-DI
07340058410868
PMA / PMN Number
K091657
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON CONTACT WITH (B)(4), THE AGENCY INFORMED EURO DIAGNOSTICA THAT THE IMMUNOSCAN CCPLUS KITS WERE PURCHASED, IN TOTAL (B)(4) KITS OVER THE LAST 3-4 MONTHS, FROM (B)(4). THE KITS WERE ONLY SUPPLIED TO (B)(6). (B)(4) SAID THAT NEITHER (B)(6) NOR (B)(4) HAD ANY STOCK LEFT OF THESE IMMUNOSCAN CCPLUS KITS FROM (B)(4)\. TO THE BEST OF OUR KNOWLEDGE, THIS IS THE ONLY EURO DIAGNOSTICA PRODUCT COUNTERFEIT.

Description of Event or Problem · 1

DURING A RETROSPECTIVE REVIEW, PERFORMED IN JAN 2018, OF COMPLAINTS FILED BY EURO DIAGNOSTICA IN 2015-2016 TO ASSESS THE COMPLAINTS FOR MDR AND/OR RECALL REPORTABILITY, ONE COMPLAINT (B)(4) WAS IDENTIFIED AS REQUIRING REMEDIATION. A MDR IS THEREFORE SUBMITTED TO THE FDA RETROSPECTIVELY FOR THE COMPLAINT. IT CAME TO EURO DIAGNOSTICA'S ATTENTION ON (B)(4) 2016, FROM (B)(4), AN (B)(4) DISTRIBUTOR, THAT AN EURO DIAGNOSTICA PRODUCT END-USER, (B)(6), OBTAINED IMMUNOSCAN CCPLUS KITS WITH FALSIFIED LABELING. ON (B)(4) 2016, EURO DIAGNOSTICA'S QA DEPT. WAS INFORMED OF THE MATTER, AND COULD CONFIRM THAT THE LABELS WERE NOT GENUINE EURO DIAGNOSTICA LABELS. ON THE KIT BOX, TWO LABELS HAD BEEN PLACED: - ONE LABEL, PLACED ON THE SHORT SIDE OF THE KIT BOX, STATED EURO DIAGNOSTICA AS THE MANUFACTURER AND (B)(4) AS IMPORTER, WITH KIT MANUFACTURING DATE (MFD) 2015-04-16. - THE OTHER LABEL, PLACED ON THE LONG SIDE OF THE KIT BOX, WAS VERY SIMILAR TO EURO DIAGNOSTICA'S KIT BOX LABEL FOR IMMUNOSCAN CCPLUS WITH A GENUINE IMMUNOSCAN CCPLUS KIT LOT (SS 1337) AND THE CORRECT EXPIRY DATE (2016-08-18) FOR THAT LOT. THE FALSE LABEL COULD BE RECOGNIZED ON SEVERAL ITEMS, E.G. LACKING SYMBOLS, WRONG SYMBOLS AND HAVING THE SPELLING MISTAKE OF (B)(4) AS "(B)(4)" IN THE BOTTOM LEFT CORNER. ON GENUINE EURO DIAGNOSTICA IMMUNOSCAN CCPLUS KIT BOXES, ONLY ONE LABEL IS PRESENT ON THE KIT BOX LONG SIDE. (B)(4) IS A FORMER EURO DIAGNOSTICA DISTRIBUTOR FROM (B)(4). EURO DIAGNSOTICA'S LAST IMMUNOSCAN CCPLUS SHIPMENT TO (B)(4) WAS ON LOT RS 2432 (EXP. DATE 2015-12-10), ON (B)(6) 2014. THE IMMUNOSCAN CCPLUS KIT LOT SS 1337 WAS NEVER SOLD BY EURO DIAGNOSTICA TO (B)(4). EURO DIAGNOSTICA RECEIVED INFORMATION THAT THE END-USER OBTAINED THE IMMUNOSCAN CCPLUS KITS FROM A SUBDEALER, (B)(4). UPON CONTACT, (B)(4) INFORMED EURO DIAGNOSTICA THAT THE KITS WERE PURCHASED, IN TOTAL 3-4 KITS OVER THE LAST 3-4 MONTHS, FROM (B)(4). THE KITS WERE ONLY SUPPLIED TO (B)(6). (B)(4) SAID THAT NEITHER (B)(6) NOR (B)(4) HAD ANY STOCK LEFT OF THESE IMMUNOSCAN CCPLUS KITS FROM (B)(4). (B)(4) WILL, WHEN GOING FORWARD, PURCHASE THE PRODUCT FROM EURO DIAGNOSTICA'S DISTRIBUTOR, (B)(4). EURO DIAGNOSTICA CONTACTED (B)(6) TO INFORM THEM THAT NO IMMUNOSCAN CCPLUS KITS FROM (B)(4) WERE TO BE USED IN PATIENT DIAGNOSIS. THE KIT BOX WAS NEVER RETURNED TO EURO DIAGNOSTICA. NOR WERE PICTURES OBTAINED OF THE KIT CONTENT. IT COULD THEREFORE NOT BE DETERMINED IF THE KIT WAS AN EXPIRED EURO DIAGNOSTICA KIT WITH NEW, FALSIFIED, LABELS ON IT, OR IF THE KIT COMPONENTS WERE FALSIFIED AS WELL. OUR ASSESSMENT IS THAT, MOST LIKELY, THE KIT IS FROM AN EXPIRED IMMUNOSCAN CCPLUS KIT LOT WHICH WAS RELABELED AND SOLD BY (B)(4). (B)(4)'S MANAGING DIRECTOR DENIES ANY WRONGDOING. FORMAL LETTERS HAVE BEEN SENT TO (B)(4) AND THE (B)(4) HEALTH AUTHORITY (B)(4). NO REPLY WAS OBTAINED FROM EITHER PARTY. USING EXPIRED IMMUNOSCAN CCPLUS COULD LEAD TO AN INCREASED RISK OF OBTAINING FALSE NEGATIVE TEST RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
132083 IMMUNOSCAN CCPLUS CCPLUS NHX EURO DIAGNOSTICA AB SS 1337 07340058410868

Patients

Seq Age Sex Outcome Treatment
1