FDA Adverse Event Other Summary report: N

HEARTSTART

MDR report key: 728432 · Received June 23, 2006

Report

Report Number
9610483-2006-00123
Event Type
Other
Date Received
June 23, 2006
Date of Event
November 10, 2000
Report Date
November 15, 2000
Manufacturer
LAERDAL MEDICAL, A.S.
Product Code
MKJ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE RETURNED DEVICE WAS TESTED USING A KNOWN GOOD POWER SOURCE AND PROPER OPERATION WAS VERIFIED. THE 2 RETURNED EXPIRED BATTERIES WERE TESTED AS FOLLOWS: THE LAERDAL BATTERY LOT 960913 DELIVERED 4 SHOCKS BEFORE THE UNIT SHUT DOWN PROMPTING "REPLACE BATTERY, BATTERY LOW", AND THE NON-LAERDAL BATTERY LOT 940412 COULD NOT DELIVER EVEN ONE SHOCK BEFORE SUTTING DOWN THE UNIT PROMTING "REPLACED BATTERY, BATTERY LOW." BOTH BATTERIES WOULD FALL THE BATTERY CAPACITY TEST AS DEFINED IN THE HS3000 OPERATING INSTRUCTIONS. THE MESSAGE "REPLACE BATTERY, BATTERY LOW" PROMPTS WHEN THE BATTERY LACKS THE CAPACITY TO PERFORM AS REQUIRED AND THE DEFIBILLATOR SHUTS DOWN. THE INCIDENT REPORT PRINTED FROM THE RETURNED MCM DOCUMENTS 2 CYCLES OF ANALYSIS, START CHARGING, "NO SHOCK INDICATED", AND SHOCK ABORTED. THE UNIT WAS TURNED OFF AND BACK ON AGAIN WITH A "REPLACE BATTERY, BATTERY LOW MESSAGE AND SHUT DOWN. A THIRD POWER-ON RESULTED IN ANOTHER ABORTED CHARGE AND "NO SHOCK INDICATED". PER THE HS3000 OPERATING INSTRUCTIONS, THE "NO SHOCK INDICATED", MESSAGE IS DISPLAYED IN THE SEMI-AUTO MODE WHEN ECG ANALYSIS DOES NOT DETECT A SHOCKABLE RHYTHM AFTER ANALYZE IS ACTIVATED. THIS REPORT IS THE RESULT OF A RETROSPECTIVE COMPLAINT REVIEW BY LMC. IT IS TIMELY FILED UNDER LMC'S REVISED MDR PROCEDURE AND ITS WRITTEN COMMITMENT TO THE AGENCY, EACH OF WHICH HAS RESULTED FROM LAERDAL'S RECEIVING INFORMATION RELATING THE AGENCY'S CURRENT THINKING WITH RESPECT TO MDR REGULATION INTERPRETATION AND ENFORCEMENT POLICY.

Description of Event or Problem · 1

DURING AN INCIDENT ON 11/10/00 INVOLVING A FEMALE PT IN CARDIAC ARREST DURING DIALYSIS TREATMENT, THIS DEFIBRILLATOR CHARGED BUT DUMPED THE CHARGE HALFWAY THROUGH 3 TIMES. CPR WAS CONTINUED. THE CREW REPLACED THE BATTERY, BUT THAT ONE WAS DEAD SO THEY REINSTALLED THE FIRST BATTERY AND AGAIN THE UNIT DUMPED THE CHARGE. ALS ARRIVED, BEGAN ALS INTERVENTION, AND TRANSPORTED THE PT TO A HOSPITAL WITH NO CHANGE TO HER CONDITION. THE PT WAS 2 HOURS INTO A 3 HOUR DIALYSIS TREATMENT AND WAS FOUND RECLINED IN A CHAIR NOT BREATHING AND WITHOUT PULSE. THE DIALYSIS CENTER STAFF WAS PERFORMING INADEQUATE CPR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSTART SAED MKJ LAERDAL MEDICAL, A.S. HS3000 9605

Patients

Seq Age Sex Outcome Treatment
1 72 YR Other