HEARTSTART
Report
- Report Number
- 9610483-2006-00123
- Event Type
- Other
- Date Received
- June 23, 2006
- Date of Event
- November 10, 2000
- Report Date
- November 15, 2000
- Manufacturer
- LAERDAL MEDICAL, A.S.
- Product Code
- MKJ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
THE RETURNED DEVICE WAS TESTED USING A KNOWN GOOD POWER SOURCE AND PROPER OPERATION WAS VERIFIED. THE 2 RETURNED EXPIRED BATTERIES WERE TESTED AS FOLLOWS: THE LAERDAL BATTERY LOT 960913 DELIVERED 4 SHOCKS BEFORE THE UNIT SHUT DOWN PROMPTING "REPLACE BATTERY, BATTERY LOW", AND THE NON-LAERDAL BATTERY LOT 940412 COULD NOT DELIVER EVEN ONE SHOCK BEFORE SUTTING DOWN THE UNIT PROMTING "REPLACED BATTERY, BATTERY LOW." BOTH BATTERIES WOULD FALL THE BATTERY CAPACITY TEST AS DEFINED IN THE HS3000 OPERATING INSTRUCTIONS. THE MESSAGE "REPLACE BATTERY, BATTERY LOW" PROMPTS WHEN THE BATTERY LACKS THE CAPACITY TO PERFORM AS REQUIRED AND THE DEFIBILLATOR SHUTS DOWN. THE INCIDENT REPORT PRINTED FROM THE RETURNED MCM DOCUMENTS 2 CYCLES OF ANALYSIS, START CHARGING, "NO SHOCK INDICATED", AND SHOCK ABORTED. THE UNIT WAS TURNED OFF AND BACK ON AGAIN WITH A "REPLACE BATTERY, BATTERY LOW MESSAGE AND SHUT DOWN. A THIRD POWER-ON RESULTED IN ANOTHER ABORTED CHARGE AND "NO SHOCK INDICATED". PER THE HS3000 OPERATING INSTRUCTIONS, THE "NO SHOCK INDICATED", MESSAGE IS DISPLAYED IN THE SEMI-AUTO MODE WHEN ECG ANALYSIS DOES NOT DETECT A SHOCKABLE RHYTHM AFTER ANALYZE IS ACTIVATED. THIS REPORT IS THE RESULT OF A RETROSPECTIVE COMPLAINT REVIEW BY LMC. IT IS TIMELY FILED UNDER LMC'S REVISED MDR PROCEDURE AND ITS WRITTEN COMMITMENT TO THE AGENCY, EACH OF WHICH HAS RESULTED FROM LAERDAL'S RECEIVING INFORMATION RELATING THE AGENCY'S CURRENT THINKING WITH RESPECT TO MDR REGULATION INTERPRETATION AND ENFORCEMENT POLICY.
DURING AN INCIDENT ON 11/10/00 INVOLVING A FEMALE PT IN CARDIAC ARREST DURING DIALYSIS TREATMENT, THIS DEFIBRILLATOR CHARGED BUT DUMPED THE CHARGE HALFWAY THROUGH 3 TIMES. CPR WAS CONTINUED. THE CREW REPLACED THE BATTERY, BUT THAT ONE WAS DEAD SO THEY REINSTALLED THE FIRST BATTERY AND AGAIN THE UNIT DUMPED THE CHARGE. ALS ARRIVED, BEGAN ALS INTERVENTION, AND TRANSPORTED THE PT TO A HOSPITAL WITH NO CHANGE TO HER CONDITION. THE PT WAS 2 HOURS INTO A 3 HOUR DIALYSIS TREATMENT AND WAS FOUND RECLINED IN A CHAIR NOT BREATHING AND WITHOUT PULSE. THE DIALYSIS CENTER STAFF WAS PERFORMING INADEQUATE CPR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTSTART | SAED | MKJ | LAERDAL MEDICAL, A.S. | HS3000 | 9605 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Other |