LMA CLASSIC, REU, SIZE 4 (100040)
Report
- Report Number
- 9681900-2018-00015
- Event Type
- Malfunction
- Date Received
- February 16, 2018
- Date of Event
- January 31, 2018
- Report Date
- February 5, 2018
- Manufacturer
- TELEFLEX MEDICAL
- Product Code
- CAE
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- NURSE
Narratives
(B)(4). THE SAMPLE WAS RETURNED FOR EVALUATION. A VISUAL EXAM WAS PERFORMED AND IT WAS OBSERVED THAT THE AIRWAY TUBE OF THE DEVICE WAS DISCOLORED DUE TO MULTIPLE USES. WHEN THE DEVICE WAS CONNECTED TO A SYRINGE, THE SYRINGE WAS UNABLE TO BLOW AIR INTO THE DEVICE OR EXTRACT AIR FROM THEM. THE RETURNED SAMPLE WAS COMPARED TO A BRAND NEW RETAINED SAMPLE. IT WAS OBSERVED THAT THE ADAPTOR CAP OF THE CHECK VALVE WAS MISSING FROM THE RETURNED SAMPLE. A DEVICE HISTORY RECORD (DHR) REVIEW WAS PERFORMED AND THERE WERE NO ISSUES FOUND THAT COULD RELATE TO THE REPORTED COMPLAINT. BASED ON THE INVESTIGATION PERFORMED, THE REPORTED COMPLAINT WAS CONFIRMED. THE CHECK VALVE COMPONENT WAS PROBABLY DETACHED/BROKEN DUE TO HANDLING/WASHING/BRUSHING AFTER MULTIPLE CYCLES OF RE-USE AFTER UNPACKING. ONCE THE COMPONENT OF THE VALVE WAS BROKEN, THE VALVE COULD NO LONGER FUNCTION NORMALLY.
CUSTOMER COMPLAINT ALLEGES "LMA CLASSIC WILL NOT DEFLATE." ALLEGED MALFUNCTION REPORTED AS OCCURRING DURING USE. IT WAS REPORTED THERE WAS NO PATIENT INJURY OR CONSEQUENCE.
(B)(4). THE DEVICE INVOLVED IN THIS COMPLAINT HAS NOT BEEN RECEIVED BY THE MANUFACTURER AT THE TIME OF THIS REPORT. THE INVESTIGATION INTO THIS COMPLAINT IS STILL IN PROGRESS.
CUSTOMER COMPLAINT ALLEGES "LMA CLASSIC WILL NOT DEFLATE." ALLEGED MALFUNCTION REPORTED AS OCCURRING DURING USE. IT WAS REPORTED THERE WAS NO PATIENT INJURY OR CONSEQUENCE.