Narratives
Additional Manufacturer Narrative · 1
NO PT INJURY OR MEDICAL INTERVENTION REQUIRED. A GAMBRO TECHNICAL SERVICES (GTS) FIELD REP INSPECTED THE MACHINE. HE FOUND A FLOW BALANCE ERROR DURING SETUP AND DURING THE SIMULATED TREATMENT. HE REPLACED AND CALIBRATED D1/D2 FLOWMETERS, CALIBRATED ALSO THE P2, AND THE UF BURRETTE. HE PERFORMED A T1 TEST, A SETUP AND A PT SIMULATION WITHOUT ALARMS OR ERRORS. CLINICAL INVESTIGATION IS ONGOING. 510(K)# IS K001156