PROGAV SHUNTSYSTEM
Report
- Report Number
- 3004721439-2018-00017
- Event Type
- Malfunction
- Date Received
- February 15, 2018
- Date of Event
- January 25, 2018
- Report Date
- September 5, 2019
- Manufacturer
- CHRISTOPH MIETHKE GMBH & CO. KG
- Product Code
- JXG
- PMA / PMN Number
- K103003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
AESCULAP INC. (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF CHRISTOPH MIETHKE GMBH & CO. KG (MANUFACTURER). EXEMPTION NUMBER: E2017044. MANUFACTURING SITE EVALUATION: EVALUATION ON-GOING.
THE VALVE WAS RETURNED IN A PLASTIC CONTAINER, DRY WITHOUT LIQUID. THE VALVE WAS AT PRESSURE LEVEL 18 CMH2O SET. A VISUAL INSPECTION WAS PERFORMED BUT THERE WERE NO DEFORMITIES OBSERVED. IN ORDER TO CHECK A POSSIBLE BLOCKAGE OF THE VALVE, WE PERFORMED A PERMEABILITY TEST. THIS TEST IS AT A HYDROSTATIC PRESSURE DIFFERENCE OF APPROXIMATELY 20 - 30 CMH2O IN FLOW DIRECTION CARRIED OUT. THE TEST SHOWED THAT THE VALVE IS NOT PERMEABLE. THE VALVE WAS ADJUSTED TO STEP UP AND DOWN ALL PRESSURE LEVELS IN 5-STEP INCREMENTS. THE INVESTIGATION SHOWED THAT THE VALVE WAS EASY TO ADJUST IN ALL PRESSURE LEVELS. TO MEASURE THE BRAKE RELEASE FORCE, WE CHECKED THE VALVE WITH A BRAKE FORCE METER. HOUSING MUST BE EXERCISED TO RELEASE THE ROTOR TO THE VALVE, THROUGH THE MAGNET INTEGRATED IN THE DEVICE. THE BRAKE FUNCTION WAS POSITIVE. THE MEASURED VALUES OF THE BRAKING FORCE ARE WITHIN THE PERMISSIBLE TOLERANCE. BASED ON INVESTIGATION RESULTS, WE COULD NOT CONFIRM THE REPORTED EVENT. NO FURTHER ACTIONS ARE REQUIRED.
COUNTRY OF COMPLAINT: (B)(6). IT IS REPORTED THAT VALVE ADJUSTMENT IS NOT ADJUSTABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 118033 | PROGAV SHUNTSYSTEM | PROGAV | JXG | CHRISTOPH MIETHKE GMBH & CO. KG | FV427T | 20029372 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |