FDA Adverse Event Malfunction Summary report: N

ULTRACISION SURG DEV, BLADES 10MM REUSABLE, OPEN

MDR report key: 727249 · Received June 19, 2006

Report

Report Number
1527736-2006-02764
Event Type
Malfunction
Date Received
June 19, 2006
Report Date
April 21, 2006
Manufacturer
ETHICON ENDO SURGERY, INC.(CINCINNATI)
Product Code
GEO
PMA / PMN Number
k990362
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

BASED ON ANALYSIS RESULTS, THIS COMPLAINT IS NOW DETERMINED TO BE A MDR MALFUNCTION. THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED. THE DEVICE WAS TESTED WITH A GENERATOR AND AN ERROR CODE 5 WAS RECEIVED. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FROM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. THEREFORE, THE INSTRUCTIONAL INSERT STATES: "SCRATCHES ON THE BLADE MAY LEAD TO PREMATURE BLADE FAILURE" AND "AVOID ACCIDENTAL CONTACT WITH OTHER INSTRUMENTS DURING USE." THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MFG PROCESS. 510(K) NUMBER IS K990362.

Description of Event or Problem · 1

DURING A CHOLECYSTECTOMY PROCEDURE NO FUNCTION AFTER A FEW MINUTES. A NEW DEVICE WAS USED TO COMPLETE THE CASE WITH NO ADVERSE PT CONSEQUENCE. THERE IS NO FURTHER INFO AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION SURG DEV, BLADES 10MM REUSABLE, OPEN GEO GEO ETHICON ENDO SURGERY, INC.(CINCINNATI) NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN 1.GENERATOR, 2.HAND PIECE.