FDA Adverse Event Malfunction Summary report: N

PELVIC ARRAY ASSY

MDR report key: 7272014 · Received February 14, 2018

Report

Report Number
3005985723-2018-00075
Event Type
Malfunction
Date Received
February 14, 2018
Date of Event
January 17, 2018
Report Date
June 19, 2018
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
PMA / PMN Number
K170593
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

REPORTED EVENT: IT WAS REPORTED THAT THE PELVIC ARRAY ASSEMBLY BROKE WHILE TIGHTENING. PRODUCT EVALUATION AND RESULTS: VISUAL INSPECTION SHOWS THUMB SCREW DAMAGE. SEE ATTACHMENT 1 FOR IMAGE OF THE INSTRUMENT. DIMENSIONAL INSPECTION WAS NOT COMPLETED SINCE VISUAL AND FUNCTIONAL INSPECTION CONFIRMED DAMAGED THUMB SCREW. FUNCTIONAL INSPECTION CONFIRMED PELVIC ARRAY ASSEMBLY, P/N 112230 DAMAGED THUMB SCREW NOT ENGAGING PROPERLY. PRODUCT HISTORY REVIEW: REVIEW OF THE PRODUCT HISTORY RECORDS INDICATE 50 DEVICES WERE MANUFACTURED UNDER LOT NO 1945616 AND ACCEPTED INTO FINAL STOCK ON 10/01/2016. NO NON-CONFORMANCES WERE IDENTIFIED DURING INSPECTION. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED TO P/N 112230, LOT 19450616 SHOWS NO ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. CONCLUSIONS: VISUAL INSPECTION CONFIRMED DAMAGED THUMB SCREW. FUNCTIONAL INSPECTION CONFIRMED THUMB SCREW NOT ENGAGING ONTO ARRAY ADAPTER. CORRECTIVE ACTION/PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD.

Description of Event or Problem · 0

BROKEN PELVIC ARRAY. THA PROCEDURE DELAYED FOR 5 MINUTES.

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

BROKEN PELVIC ARRAY. THA PROCEDURE DELAYED FOR 5 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
113672 PELVIC ARRAY ASSY STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. 19450616

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization