FDA Adverse Event
Other
Summary report: N
CASTLE
MDR report key: 726589
·
Received June 13, 2006
Report
- Report Number
- 3019090-2006-00001
- Event Type
- Other
- Date Received
- June 13, 2006
- Report Date
- April 6, 2006
- Manufacturer
- GETINGE USA, INC-F.K.A GETINGE/CASTLE
- Product Code
- FSY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RP
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE BREAK MAY HAVE BEEN DUE TO METAL FATIGUE. PART WAS NOT RETURNED FOR EVALUATION. A NEW YOKE HAS BEEN INSTALLED, FUNCTIONS TESTED AND RETURNED TO END USER.
Description of Event or Problem · 1
PATIENT WAS SEDATED AND ANESTHESIOLOGIST WAS IN THE PROCESS FO POSITIONING THE LIGHT WHEN THE YOKE BROKE OFF AND FELL. THE LIGHTHEAD MISSED THE PATIENT AND WAS DANGLING BESIDE THE SUGICAL TABLE SUPPORTED BY ITS POWER SUPPLY CABLES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CASTLE | SURGICAL LIGHT | FSY | GETINGE USA, INC-F.K.A GETINGE/CASTLE | GCL 550 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |