FDA Adverse Event Injury Summary report: N

TITAN OTR NAR SCRO 16CM

MDR report key: 7264565 · Received February 12, 2018

Report

Report Number
2125050-2018-00100
Event Type
Injury
Date Received
February 12, 2018
Date of Event
January 21, 2018
Report Date
June 6, 2018
Manufacturer
COLOPLAST A/S
Product Code
FHW
PMA / PMN Number
P000006
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

WITHOUT THE BENEFIT OF EXAMINATION AND TESTING, COLOPLAST IS PRECLUDED FROM COMMENTING ON THE CONDITION OF THE DEVICE OR THE CAUSE OF THE OCCURRENCE.  SHOULD ADDITIONAL FACTS PROMPT US TO ALTER OR SUPPLEMENT ANY INFORMATION OF CONCLUSIONS CONTAINED IN THE ORIGINAL MDR OR IN ANY PRIOR SUPPLEMENTAL REPORTS, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP MDR IS CREATED TO DOCUMENT THE CONCLUSION OF THE INVESTIGATION. A TITAN OTR PUMP, TWO CYLINDERS, AND A RESERVOIR WERE RECEIVED FOR EVALUATION. HOWEVER, BECAUSE QUALITY'S EXAMINATION MAY NOT CONCLUSIVELY CONFIRM OR DISPROVE THE REPORT OF EROSION AND EXPOSURE TO GLANS, QUALITY ACCEPTS THE PHYSICIAN'S OBSERVATIONS OF SUCH AS THE REASON FOR SURGICAL INTERVENTION. QUALITY REVIEWED THE MANUFACTURING RECORDS FOR THIS DEVICE AND CONFIRMED THAT THERE WERE NO DISCREPANCIES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT AND VERIFIED THAT THE DEVICES FROM THIS LOT MET ALL SPECIFICATIONS PRIOR TO RELEASE. A REVIEW OF THE COMPLAINT HISTORY DATABASE REVEALED NO ADDITIONAL COMPLAINTS REPORTED FOR LOT 4241670. REVIEW OF NONCONFORMING REPORTS REVEALED NO NONCONFORMANCE'S WITH LOT 4241670 THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. NO CAPAS FOR ISSUES OF THIS TYPE ARE ASSOCIATED WITH LOT 4241670. MANAGEMENT ROUTINELY REVIEWS EVENTS SUCH AS THIS AND MONITORS COMPLAINT LEVELS. ADDITIONALLY, EVENTS OF THIS TYPE ARE CAPTURED IN THE PRODUCT RISK DOCUMENTATION. BASED ON THIS INFORMATION, NO FURTHER CORRECTIVE ACTION IS REQUIRED AT THIS TIME.

Description of Event or Problem · 1

ACCORDING TO AVAILABLE INFORMATION, EROSION AND CYLINDER EXPOSURE TO GLANS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
109059 TITAN OTR NAR SCRO 16CM INFLATABLE PENILE PROSTHESIS FHW COLOPLAST A/S QNR9161022 4241670

Patients

Seq Age Sex Outcome Treatment
1 62 YR Other