DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2018-014848
- Event Type
- Malfunction
- Date Received
- February 9, 2018
- Date of Event
- January 11, 2018
- Report Date
- January 12, 2018
- Manufacturer
- DEXCOM, INC.
- Product Code
- PQF
- UDI-DI
- 10386270000221
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE PATIENT EXPERIENCED ADVERSE EVENTS ON DIFFERENT DAYS WITH THE SAME DEVICE. THE FOLLOWING REPORTS ARE BEING SUBMITTED: 180209-001084.
DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018, A LOSS OF CONNECTION BETWEEN THE TRANSMITTER AND SMART DEVICE OCCURRED. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THERE WAS NO FAILURE DETECTED RELATED TO THE COMPLAINT. A PAIRING TEST WAS PERFORMED WITH A KNOWN GOOD TRANSMITTER AND PASSED. VOLTAGE TESTING WAS PERFORMED AND PASSED. A REVIEW OF THE DOWNLOADED DATA LOG CONFIRMED THE REPORTED EVENT OF LOSS OF CONNECTION. A ROOT CAUSE COULD NOT BE DETERMINED.