FDA Adverse Event Malfunction Summary report: N

ENDOPATH ETS FLEX45 ENDOSCOPIC ARTICULATING LINEAR CUTTER

MDR report key: 726067 · Received June 8, 2006

Report

Report Number
1527736-2006-02610
Event Type
Malfunction
Date Received
June 8, 2006
Date of Event
June 15, 2005
Report Date
June 20, 2005
Manufacturer
ETHICON ENDO-SURGERY, INC.(CINCINNATI)
Product Code
GDW
PMA / PMN Number
k020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: BASED ON THE ANALYSIS FINDINS OF DAMAGED PINION AXLE AND DAMAGED FIRING TRIGGER TEETH, THIS EVENT IS CONSIDERED TO BE REPORTABLE. THE DEVICE WAS RECEIVED WITH THE FIRING MECHANISM DAMAGED AND WITH A CARTRIDGE LOADED IN THE DEVICE. THE CARTRIDGE WAS RECEIVED PARTIALLY FIRED AND WITH NO DAMAGE TO THE CARTRIDGE LOCK OUT TAB. THE RETURNED DEVICE WAS FOUND TO BE NON FUNCTIONAL AS THE FIRING MECHANISM WAS NOTED TO BE DAMAGED. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS; THE LEFT SHROUD PINION AXLE SUPPORT WAS FOUND DAMAGED AND THE FIRING TRIGGER TEETH WERE FOUND BROKEN. NO FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE CONDITION OF THE DEVICE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS. 510(K) IS K020779

Description of Event or Problem · 1

DURING A LAPAROSCOPIC GASTRIC BYPASS PROCEDURE, WHEN HE WAS MAKING POUCH AFTER ABOUT 4 FIRINGS ON THE STOMACH HE NOTICED THAT THE STAPLER FELT FUNNY WHEN HE FIRED IT. WHEN HE CHECKED THE STAPLE LINE HE SAW THAT THE STAPLES FIRED BUT DID NOT CUT. HE THEN FIRED AGAIN WITH ANOTHER GUN AND THE SAM3E THING HAPPENED ANOTHER SHORT FIRE. HE THEN OPENED ANOTHER GUN AND FIRED AGAIN, THIS TIME THE STAPLER WORKED FINE. THERE WAS NO PT CONSEQUENCES. THE CASE WAS COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH ETS FLEX45 ENDOSCOPIC ARTICULATING LINEAR CUTTER GDW GDW ETHICON ENDO-SURGERY, INC.(CINCINNATI) NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN