ULTRACISION SURG DEV, CURVED SHEARS, ENDO
Report
- Report Number
- 1527736-2006-02599
- Event Type
- Malfunction
- Date Received
- June 7, 2006
- Report Date
- April 21, 2006
- Manufacturer
- ETHICON ENDO SURGERY, INC.(CINCINNATI)
- Product Code
- LFL
- PMA / PMN Number
- k002981
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED. DUE TO THE DAMAGE TO THE BLADE, IT WAS CONFIRMED THAT THE DEVICE WAS NON-FUNCTIONAL. AN ERROR CODE 5 WAS NOTED. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FORM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. THEREFORE, THE INSTRUCTIONAL INSERT STATES: "SCRATCHES ON THE BLADE MAY LEAD TO PREMATURE BLADE FAILURE" AND "AVOID ACCIDENTAL CONTACT WITH OTHER INSTRUMENTS DURING USE." (510(K)#K002981)
DURING A LAP SIGMA PROCEDURE, AFTER 10 MINUTES, ERROR CODE. TOOK NEW INSTRUMENT. NO FURTHER INFORMATION IS AVAILABLE. THERE WAS NO ADVERSE PATIENT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ULTRACISION SURG DEV, CURVED SHEARS, ENDO | GEI | LFL | ETHICON ENDO SURGERY, INC.(CINCINNATI) | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | GENERATOR| HANDPIECE |