FDA Adverse Event Malfunction Summary report: N

ULTRACISION SURG DEV, BLADES 10MM REUSABLE, OPEN

MDR report key: 725969 · Received June 7, 2006

Report

Report Number
1527736-2006-02597
Event Type
Malfunction
Date Received
June 7, 2006
Report Date
April 21, 2006
Manufacturer
ETHICON ENDO SURGERY, INC.(CINCINNATI)
Product Code
KNS
PMA / PMN Number
k990362
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH THE BLADE SCRATCHED, NICKED AND CRACKED. DUE TO THE DAMAGE TO THE BLADE, IT WAS CONFIRMED THAT THE DEVICE WAS NON-FUNCTIONAL. AN ERROR CODE 5 WAS NOTED. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FROM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. THEREFORE, THE INSTRUCTIONAL INSET STATES: "SCRATCHES ON THE BLADE MAY LEAD TO PREMATURE BLADE FAILURE" AND "AVOID ACCIDENTAL CONTACT WITH OTHER INSTRUMENTS DURING USE." (510(K)#K990362)

Description of Event or Problem · 1

DURING A CHOLECYSTECTOMY PROCEDURE, AFTER 5 TIMES, NO FUNCTION. DISPLAY SHOWS INSTRUMENT ERROR. TOOK NEW INSTRUMENT. NO FURTHER INFORMATION IS AVAILABLE. THERE WAS NO ADVERSE PATIENT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION SURG DEV, BLADES 10MM REUSABLE, OPEN GEI KNS ETHICON ENDO SURGERY, INC.(CINCINNATI) NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN HANDPIECE| GENERATOR