ULTRACISION SURG DEV, BLADES 10MM REUSABLE, OPEN
Report
- Report Number
- 1527736-2006-02597
- Event Type
- Malfunction
- Date Received
- June 7, 2006
- Report Date
- April 21, 2006
- Manufacturer
- ETHICON ENDO SURGERY, INC.(CINCINNATI)
- Product Code
- KNS
- PMA / PMN Number
- k990362
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH THE BLADE SCRATCHED, NICKED AND CRACKED. DUE TO THE DAMAGE TO THE BLADE, IT WAS CONFIRMED THAT THE DEVICE WAS NON-FUNCTIONAL. AN ERROR CODE 5 WAS NOTED. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FROM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. THEREFORE, THE INSTRUCTIONAL INSET STATES: "SCRATCHES ON THE BLADE MAY LEAD TO PREMATURE BLADE FAILURE" AND "AVOID ACCIDENTAL CONTACT WITH OTHER INSTRUMENTS DURING USE." (510(K)#K990362)
DURING A CHOLECYSTECTOMY PROCEDURE, AFTER 5 TIMES, NO FUNCTION. DISPLAY SHOWS INSTRUMENT ERROR. TOOK NEW INSTRUMENT. NO FURTHER INFORMATION IS AVAILABLE. THERE WAS NO ADVERSE PATIENT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ULTRACISION SURG DEV, BLADES 10MM REUSABLE, OPEN | GEI | KNS | ETHICON ENDO SURGERY, INC.(CINCINNATI) | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | HANDPIECE| GENERATOR |