ENDOPATH ETS45 ENDOSCOPIC LINEAR CUTTER
Report
- Report Number
- 1527736-2006-02618
- Event Type
- Malfunction
- Date Received
- June 8, 2006
- Date of Event
- March 27, 2006
- Report Date
- March 28, 2006
- Manufacturer
- ETHICON ENDO-SURGERY, INC.(CINCINNATI)
- Product Code
- KOG
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PHYSICIAN
Narratives
EVALUATION SUMMARY: BASED ON THE ANALYSIS FINDING OF DAMAGED FIRING TRIGGER TEETH, THIS FILE IS CONSIDERED TO BE REPORTABLE. THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A CARTRIDGE LOADED IN THE DEVICE. THE CARTRIDGE WAS RECEIVED PARTIALLY FIRED AND WITH THE CARTRIDGE LOCK OUT TAB BENT. THE DAMAGE TO THE CARTRIDGE LOCKOUT TAB IS CONSISTENT WITH DAMAGE OBSERVED WHEN THE FIRING CYCLE IS STARTED, INTERRUPTED, RELEASED, AND RESTARTED. IN ADDITION, THE INSTRUCTIONS FOR USE STATE, " THE FIRING STROKE MUST BE COMPLETED. DO NOT PARTIALLY FIRE THE DEVICE. FIRE THE DEVICE BY SQUEEZING THE FIRING TRIGGER COMPLETELY UNTIL IT RESTS ON THE CLOSING TRIGGER. NOTE: ONCE THE FIRING CYCLE HAS BEEN INITIATED, IT MUST BE COMPLETED. IF RE-INITIATION OF FIRING IS RESUMED, THE DEVICE WILL LOCKOUT. FING THROUGH THE LOCKOUT MECHANISM WILL BREAK THE DEVICE." THE RETURNED DEVICE WAS FOUND TO BE NON FUNCTIONAL AS THE FIRING MECHANISM WAS NOTED TO BE DAMAGED. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE FIRING TRIGGER TEETH WERE FOUND BROKEN. NO FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE CONDITION OF THE DEVICE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS. 510(K) IS K020779
DURING A GYN PROCEDURE, THE DEVICE WOULD NOT CLOSE IT WAS NOT USED, PULLED NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH ETS45 ENDOSCOPIC LINEAR CUTTER | GDW | KOG | ETHICON ENDO-SURGERY, INC.(CINCINNATI) | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN |