INTROCAN SAFETY
Report
- Report Number
- 2523676-2006-00046
- Event Type
- Other
- Date Received
- June 7, 2006
- Date of Event
- April 12, 2006
- Report Date
- May 24, 2006
- Manufacturer
- B. BRAUN MEDICAL, INC.
- Product Code
- FOZ
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE ACTUAL DEVICE IN THE INCIDENT WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. THE SAMPLE CONSISTED OF THE CATHETER HUB WITHOUT THE CATHETER ATTACHED TO THE SPINLOCK CONNECTOR OF AN UNKNOWN IV SET. THE DETACHED CATHETER WAS ALSO RETURNED, LYING LOOSE IN THE BAG. THE INTERNAL STAINLESS STEEL FUNNEL USED TO LOCK THE CATHETER INTO THE HUB COULD CLEARLY BE SEEN AND WAS INTACT. NO PORTION OF THE CATHETER REMAINED INSIDE THE HUB. THE FRACTURED CATHETER APPEARED INTACT. ALTHOUGH THERE IS NO CURRENT, UNUSED INVENTORY OF THE REPORTED POSSIBLE LOTS OF PRODUCT, 25 UNUSED SAMPLES, FROM A LOT CURRENTLY IN B. BRAUN'S INVENTORY WERE SUBJECTED TO A PULL TEST OF THE CATHETER TO HUB JOINT UNTIL FAILURE USING A TENSILE FORCE TESTER. ALL SAMPLES EXCEEDED THE MINIMUM JOINT SEPARATION DESIGN SPECIFICATION AND PASSED THE PULL TEST. THE SAMPLE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED TO THE ACTUAL MANUFACTURER, B. BRAUN MEDICAL INDUSTRIES SDN BHD, IN MALAYSIA.
AS REPORTED BY THE USER FACILITY: WHEN REMOVING CATHETER FROM THE PATIENT THE HUB BROKE OFF. THEY WERE ABLE TO COMPLETELY REMOVE THE CATHETER BEFORE IT MIGRATED. THERE WAS NO PATIENT INJURY. ADDITIONAL INFORMATION RECEIVED FROM THE FACILITY INDICATED THAT THE CATHETER WAS REMOVED INTACT AND THE PATIENT SUFFERED NO ADVERSE SEQUELA ASSOCIATED WITH THIS INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INTROCAN SAFETY | I.V. SAFETY CATHETER | FOZ | B. BRAUN MEDICAL, INC. | NA | 5M15258W07 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR | Other |