VISERA CYSTO-NEPHRO VIDEOSCOPE
Report
- Report Number
- 8010047-2018-00143
- Event Type
- Injury
- Date Received
- February 6, 2018
- Date of Event
- December 19, 2017
- Report Date
- February 22, 2019
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORP.
- Product Code
- FAJ
- PMA / PMN Number
- K133538
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- RISK MANAGER
Narratives
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE ADDITIONAL DEVICE EVALUATION RESULT. THE ADDITIONAL EVALUATION OF THE TWO CYF-VA2 BY OMSC CONFIRMED FOLLOWING. [CYF-VA2, S/N (B)(4)]. THERE WAS A PINHOLE ON THE RUBBER OF THE BENDING SECTION. THE BENDING ANGLE DOES NOT MEET THE SPECIFICATION. THE REMOTE SWITCH 4 DID NOT WORK. THERE WERE RESIDUAL SUBSTANCES OF WATER INSIDE OF THE CONTROL SECTION, UNIVERSAL CORD AND THE LIGHT GUIDE LENS. THERE WERE AIR LEAKAGES AT THE INSTRUMENT CHANNEL AND THE RUBBER OF THE BENDING SECTION. WHEN INSERTING AN ENDOTHERAPY ACCESSORY INTO THE INSTRUMENT CHANNEL, STUCK WAS FELT AT THE BENDING SECTION AND UNDER THE BOOT. WHEN INSERTING A CHANNEL CLEANING BRUSH INTO THE INSTRUMENT CHANNEL, STUCK WAS FELT AT THE BENDING SECTION. THERE WAS A KINK AT THE BENDING SECTION. THERE WAS DEBRIS AROUND THE INSTRUMENT CHANNEL PORT. [CYF-VA2, S/N (B)(4)]. THERE WAS DEBRIS INSIDE OF THE INSTRUMENT CHANNEL. THE EXACT CAUSE COULD NOT BE DETERMINED. THIS REPORT IS 1 OF 8 REPORTS.
THIS SUPPLEMENTAL REPORT IS SUBMITTING TO CORRECT "DEVICE PRODUCT CODE" AND "PMA/510(K) NUMBER".
THE TWO CYF-VA2 WERE RETURNED TO OMSC FOR EVALUATION. THE EVALUATION FOUND FOLLOWING FAILURES. [CYF-VA2, S/N (B)(4), MANUFACTURED ON SEPTEMBER 13, 2015] - THERE WERE KINK IN THE INSTRUMENT CHANNEL AND THE INSERTION TUBE. - THERE WERE DEBRIS AND SCRATCHES INSIDE OF THE INSTRUMENT CHANNEL. - THERE WERE SCRATCHES ON THE RUBBER OF THE BENDING SECTION. - THE PART OF THE COATING OVER THE INSERTION TUBE WAS PEELING OFF. [CYF-VA2, S/N (B)(4), MANUFACTURED ON NOVEMBER 11, 2010] - THERE WERE KINK IN THE INSTRUMENT CHANNEL AND THE INSERTION TUBE, THE UNIVERSAL CORD AND THE INSTRUMENT CHANNEL. - THERE WERE SCRATCHES INSIDE OF THE INSTRUMENT CHANNEL. - THERE WERE SCRATCHES ON THE RUBBER OF THE BENDING SECTION AND THE ADHESIVE OF BENDING RUBBER. - THERE WERE DEBRIS ON THE UNIVERSAL CORD. OMSC REVIEWED THE MANUFACTURE HISTORY OF THE TWO CYF-VA2 AND CONFIRMED NO IRREGULARITY. THE EXACT CAUSE COULD NOT BE DETERMINED AT PRESENT, IF SIGNIFICANT ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT 8 PATIENTS HAD A FEVER AFTER A CYSTOSCOPY USING SUBJECT DEVICE. THE 8 PATIENTS REPORTEDLY EXPERIENCED CYSTOSCOPY BETWEEN DECEMBER 2017 AND JANUARY 2018. THE FACILITY PROVIDED UNSPECIFIED MEDICAL TREATMENT FOR THE PATIENTS, BUT THERE WAS NO INFORMATION OF THE OUTCOME. THE FACILITY POSSESSES TWO CYF-VA2 (SERIAL (B)(4)), BUT IT IS UNKNOWN WHICH SCOPE WAS USED FOR EACH PATIENT. THE USER FACILITY CONDUCTED MICROBIAL CULTURE TEST FOR TWO CYF-VA2 AND THE TESTING INDICATED NO MICROBIAL GROWTH FOR THE TWO CYSTOSCOPES. THE USER FACILITY REPORTED THAT THE SUBJECT DEVICE HAD BEEN REPROCESSED USING ENDOSTREAM, A NON-OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR MODEL, THIS REPORT IS 1 OF 8 REPORTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 88953 | VISERA CYSTO-NEPHRO VIDEOSCOPE | CYSTO-NEPHRO VIDEOSCOPE | FAJ | OLYMPUS MEDICAL SYSTEMS CORP. | CYF-VA2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |