FDA Adverse Event Malfunction Summary report: N

VITEK® 2 GP TEST KIT

MDR report key: 7234374 · Received February 1, 2018

Report

Report Number
1950204-2018-00053
Event Type
Malfunction
Date Received
February 1, 2018
Report Date
March 20, 2018
Manufacturer
BIOMERIEUX, INC
Product Code
LQL
PMA / PMN Number
C1 EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A CUSTOMER FROM (B)(6) REPORTED MISIDENTIFICATIONS OF ENTEROCOCCUS FAECALIS AS ENTEROCOCCUS FAECIUM FOR THREE (3) DIFFERENT PATIENT ISOLATES IN ASSOCIATION WITH THE VITEK® 2 GP TEST KIT ( LOT 2420509103). AN INVESTIGATION WAS PERFORMED. THE CUSTOMER REPORTED SETTING UP THE STRAINS FROM 18-24 HOUR CULTURES THAT WERE GROWN ON COS MEDIA. PATIENT 1: TWO (2) LAB REPORTS WERE SUBMITTED WITH BOTH SHOWING VERY GOOD IDENTIFICATIONS OF E. FAECIUM. ONE LAB REPORT HAD FOUR (4) ATYPICAL NEGATIVE REACTIONS (LEUA, TYRA, ASPA, AGLU) AND THE OTHER LAB REPORT HAD FIVE (5) ATYPICAL NEGATIVE REACTIONS (LEUA, TYRA, ASPA, AGLU, NOVO) FOR AN IDENTIFICATION OF E. FAECALIS ACCORDING TO THE GP KNOWLEDGE BASE. PATIENT 2: ONE (1) LAB REPORT WAS SUBMITTED SHOWING A VERY GOOD IDENTIFICATION OF E. FAECIUM WITH ONE (1) ATYPICAL NEGATIVE REACTION (LEUA) FOR AN IDENTIFICATION OF E. FAECALIS ACCORDING TO THE GP KNOWLEDGE BASE. PATIENT 3: ONE (1) LAB REPORT WAS SUBMITTED SHOWING A VERY GOOD IDENTIFICATION OF E. FAECIUM WITH THREE (3) ATYPICAL NEGATIVE REACTIONS (LEUA, ASPA, AGLU) FOR AN IDENTIFICATION OF E. FAECALIS ACCORDING TO THE GP KNOWLEDGE BASE. A SECOND LAB REPORT WAS SUBMITTED SHOWING A LOW DISCRIMINATION CALL BETWEEN E. FAECIUM AND E. FAECALIS. AN INCREASED NUMBER OF ATYPICAL NEGATIVE RESULTS CAN INDICATE A STRAIN WITH DECREASED VIABILITY, USER SET UP ERROR OR AN ATYPICAL STRAIN. THE THREE (3) STRAINS WERE SUBMITTED BY THE CUSTOMER FOR EVALUATION. THE SECOND STRAIN (S2) WAS CONTAMINATED BY MOLD. TESTS WERE CONDUCTED ONLY WITH S1 AND S3. THE INTENDED IDENTIFICATION TO E. FAECALIS WAS CONFIRMED ON VITEK MS V3 (KNOWLEDGE BASE V3.0). ON VITEK 2 (V7.01), ONE (1) GP CARD OF THE CUSTOMER LOT (CL : 2420509103) AND ONE (1) GP CARD OF A RANDOM LOT (RL : 2420490403) WERE TESTED FROM CBA SUBCULTURE. THESE TESTS GAVE AN EXCELLENT IDENTIFICATION TO E. FAECALIS 99% WITH BOTH STRAINS AND BOTH LOTS. THE CUSTOMER MISIDENTIFICATION TO E. FAECIUM WAS NOT REPRODUCED IN-HOUSE WHATEVER THE LOT TESTED. THE VITEK 2 GP CARD PERFORMED AS INTENDED AND NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

A CUSTOMER FROM (B)(6) CUSTOMER REPORTED MISIDENTIFICATIONS FOR THREE (3) DIFFERENT PATIENT ISOLATES IN ASSOCIATION WITH THE VITEK® 2 GP TEST KIT ( LOT 2420509103). THE CUSTOMER STATED THAT FOR THE THREE (3) ISOLATES, THE IDENTIFICATION WAS ENTEROCOCCUS FAECIUM (93%) WITH THE GP CARD, AND ENTEROCCOCUS FAECALIS WHEN TESTING WITH VITEK® MS. THE CUSTOMER STATED THE MEDIA USED WAS COS, AND THE QUALITY AND PURITY CONTROLS CONFORMED. THERE WAS A DELAY GREATER THAN 24 HOURS FOR REPORTING RESULTS. THERE IS NO INDICATION OR REPORT FROM THE HOSPITAL OR TREATING PHYSICIAN TO BIOMÉRIEUX THAT THE DISCREPANT RESULT LED TO ANY ADVERSE EVENT RELATED TO THE PATIENT'S STATE OF HEALTH. A BIOMÉRIEUX INTERNAL INVESTIGATION WILL BE INITIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
78082 VITEK® 2 GP TEST KIT VITEK® 2 GP TEST KIT LQL BIOMERIEUX, INC 2420509103

Patients

Seq Age Sex Outcome Treatment
1