FDA Adverse Event Malfunction Summary report: N

ACCUVEIN

MDR report key: 7230566 · Received January 30, 2018

Report

Report Number
MW5074984
Event Type
Malfunction
Date Received
January 30, 2018
Report Date
January 29, 2018
Manufacturer
ACCUVEIN
Product Code
KZA
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

THE ACCUVEIN HF470 HANDS-FREE ROLLING STAND FOR THE ACCUVEIN AV400 REGULARLY BREAKS AT THE PROXIMAL END OF THE EXTENSION ARM. THE ARM COMPLETELY BREAKS OFF THE STAND. THE BREAKS OCCUR IN USE AND DURING TRANSPORT. WE BELIEVE ONE OF THE FACTORS TO THE BREAKS IS THAT THE END USER HAS TO APPLY REGULAR DOWNWARD FORCE TO LOCK THE ARM IN AN ARM-LOCK MECHANISM THAT IS LOCATED CLOSE TO THE BREAK. THIS REGULAR STRESS CAUSES A SMALL CRACK TO APPEAR WITH REGULAR USE, WHICH THEN GROWS, AND EVENTUALLY RESULTS IN A COMPLETE BREAK. SOMETIMES THE BREAK OCCURS INTERNALLY WITHOUT THE ARM BREAKING COMPLETELY OFF, WHICH DAMAGES A POWER CORD AND PREVENTS THE AV400 FROM CHARGING WHILE IN THE STAND. UNFORTUNATELY, WE DO NOT HAVE GOOD SOLID NUMBERS ON THE NUMBER OF BROKEN STANDS B/C WE DO NOT TRACK THESE STANDS THE SAME WAY AS THE MEDICAL DEVICE THAT ATTACHES TO THE STAND. WE KNOW HOW MANY AV400S WE HAVE, BUT NOT EXACTLY HOW MANY STANDS (THEY DON'T ALL HAVE STANDS). HOWEVER, OUT OF THE 20-30 STANDS WE HAVE IN PLACE, WE BELIEVE ROUGHLY 10 HAVE BROKEN IN THE PAST 2 YEARS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
72628 ACCUVEIN ACCUVEIN KZA ACCUVEIN HF470
72629 ACCUVEIN ACCUVEIN KZA ACCUVEIN HF470
72630 ACCUVEIN ACCUVEIN KZA ACCUVEIN HF470
72631 ACCUVEIN ACCUVEIN KZA ACCUVEIN HF470
72632 ACCUVEIN ACCUVEIN KZA ACCUVEIN HF470
72633 ACCUVEIN ACCUVEIN KZA ACCUVEIN HF470
72634 ACCUVEIN ACCUVEIN KZA ACCUVEIN HF470
72635 ACCUVEIN ACCUVEIN KZA ACCUVEIN HF470
72636 ACCUVEIN ACCUVEIN KZA ACCUVEIN HF470
72637 ACCUVEIN ACCUVEIN KZA ACCUVEIN HF470

Patients

Seq Age Sex Outcome Treatment
1