FDA Adverse Event
Other
Summary report: N
HILAN XS PNEUMATIC MOTOR
MDR report key: 722725
·
Received May 31, 2006
Report
- Report Number
- 2916714-2006-00035
- Event Type
- Other
- Date Received
- May 31, 2006
- Date of Event
- May 8, 2006
- Report Date
- May 30, 2006
- Manufacturer
- AESCULAP, INC.
- Product Code
- HBB
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
TWO DIFFERENT LUMBAR DISCECTOMY PROCEDURES, SAME SURGEON. BOTH MOTORS AND/OR HANDPIECES HEATED UP AND ONE HEATED UP SO MUCH THAT IT BURNED THE SURGEON'S INNER ASPECT ON INDEX FINGER DURING THE SECOND PROCEDURE. NO PROLONGING OF SURGERY OR PT INCIDENT FOR EITHER CASE. NOT SURE WHICH SET WAS USED DURING PROCEDURE WHEN SURGEON WAS BURNED. SURGEON WAS DOUBLE GLOVED. HE HAD TO BE TREATED FOR 1ST DEGREE BURN (SILVADANE OINTMENT TO TREAT BURN). REFERENCE: 2916714-2006-00036 DEVICE 2.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HILAN XS PNEUMATIC MOTOR | MOTOR, DRILL, PNEUMATIC | HBB | AESCULAP, INC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN |