FDA Adverse Event Malfunction Summary report: N

CABINET, X-RAY SYSTEM

MDR report key: 7224815 · Received January 29, 2018

Report

Report Number
7224815
Event Type
Malfunction
Date Received
January 29, 2018
Date of Event
September 15, 2017
Report Date
January 17, 2018
Manufacturer
FAXITRON BIOPTICS, LLC
Product Code
MWP
Product Problem
Yes
Report Source
User Facility report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

BIOMEDICAL ENGINEERING DEPARTMENT RECEIVED A CALL THAT THE FAXITRON BIOVISION WAS LOCKED UP. IMAGING ENGINEER WENT DOWN THE OR TO INVESTIGATE. UPON ARRIVING THE IMAGING ENGINEER WITNESS THE DEVICE WAS LOCKED UP. THE ENGINEER WENT TO INVESTIGATE THE LOG FILES AND FOUND OUT THAT SOMEONE HAD DOWNLOADED MUSIC AND WAS USING THE SYSTEMS INTERNET SERVICE TO ACCESS THE MUSIC. ONCE THE DEVICE WAS RESET, THE SYSTEM WAS ABLE TO BOOT BACK UP PROPERLY AND WAS TESTED AND RETURNED TO SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
69175 CABINET, X-RAY SYSTEM MWP FAXITRON BIOPTICS, LLC BIOVISION

Patients

Seq Age Sex Outcome Treatment
1