DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2018-009716
- Event Type
- Malfunction
- Date Received
- January 26, 2018
- Date of Event
- December 28, 2017
- Report Date
- December 29, 2017
- Manufacturer
- DEXCOM, INC.
- Product Code
- PQF
- UDI-DI
- 10386270000221
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE PATIENT EXPERIENCED ADVERSE EVENTS ON DIFFERENT DAYS WITH THE SAME DEVICE. THE FOLLOWING REPORTS ARE BEING SUBMITTED: 171229-001206,171229-001233,171229-001247.
(B)(4).
DEXCOM WAS MADE AWARE ON (B)(4) 2017 THAT ON (B)(6) 2017, A LOSS OF CONNECTION BETWEEN THE TRANSMITTER AND SMART DEVICE OCCURRED. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. DATA WAS RETURNED AND EVALUATED. THE REPORTED EVENT OF A LOSS OF CONNECTION WAS CONFIRMED VIA DATA. A ROOT CAUSE COULD NOT BE DETERMINED.
THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. VOLTAGE TESTING WAS PERFORMED AND RESULTED IN 0 VOLTAGE. THERE WAS NO DATA LOG AVAILABLE TO CONFIRM THE REPORTED EVENT OF A LOSS OF CONNECTION. THE REPORTED ALLEGATION WAS NOT FOUND.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 64812 | DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM | CONTINUOUS GLUCOSE MONITOR | PQF | DEXCOM, INC. | 9438-06 | 6003458 | 10386270000221 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |