FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 7222927 · Received January 26, 2018

Report

Report Number
3008642652-2018-00999
Event Type
Malfunction
Date Received
January 26, 2018
Date of Event
December 29, 2017
Report Date
January 24, 2018
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
UDI-DI
00855778005005
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (RESETS) WAS CONFIRMED. UPON EVALUATION THE MONITOR RESET AFTER DELIVERING A PULSE DURING A PULSE TEST. THE ROOT CAUSE FOR THE RESET WAS ISOLATED TO NOISE ORIGINATING FROM THE DEFIBRILLATOR PCA HIGH-VOLTAGE CAPACITORS AND PROPAGATING ON THE MAIN BATTERY WIRE ON THE MONITOR C/A BOARD. A DESIGN CHANGE TO ADDRESS THIS CONDITION (PMA SUPPLEMENT P010030/S064) WAS APPROVED BY FDA ON (B)(4) 2015. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE MONITOR.

Description of Event or Problem · 1

A US DISTRIBUTOR REPORTED THAT A MONITOR WAS RESETTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
65649 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION WCD 4000 00855778005005

Patients

Seq Age Sex Outcome Treatment
1