PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAS
Report
- Report Number
- 3004209178-2018-46391
- Event Type
- Injury
- Date Received
- January 26, 2018
- Date of Event
- December 1, 2017
- Report Date
- May 24, 2018
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- UDI-DI
- 00643169476677
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
DEVICE RECEIVED WITH OPERATING CURRENTS WITHIN SPEC. DEVICE PASSED SELF TEST, UNEXPECTED RESTART ERROR TEST, REWIND, BASIC OCCLUSION TEST AND DISPLACEMENT TEST. HOWEVER, UNABLE TO PRIME DEVICE DURING PRIME OR COMPROMISED FORCE SENSOR SYSTEM ALARM TEST DUE TO FAULTY FORCE SENSOR RESISTOR. UNABLE TO PERFORM OCCLUSION TEST, EXCESSIVE NO DELIVERY TEST, DISPLACEMENT ACCURACY TEST OR VERIFY OVER DELIVERY ANOMALY DUE TO PRIME ANOMALY. DEVICE RECEIVED WITH CRACKED BATTERY TUBE THREADS, CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP AND MINOR SCRATCHES ON LCD WINDOW.
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE CUSTOMER WAS HOSPITALIZED DUE TO DIABETIC KETOACIDOSIS AND HIGH AND LOW BLOOD GLUCOSE TEN YEARS AGO. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 96 MG/DL AND 400 MG/DL. AT TIME OF INCIDENT AND THE BLOOD GLUCOSE LEVEL WAS 96 MG/DL. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 20 MG/DL AND 140-200 MG/DL. THE CUSTOMER WAS TREATED WITH ORANGE JUICE STOPPING TROUBLESHOOT FOR HIGH BLOOD GLUCOSE AND THE CUSTOMER WAS CURRENTLY ON LOW BLOOD GLUCOSE. THE CUSTOMER STATED THAT THE PUMP WAS NOT WORKING FOR THE PAST FEW MONTHS AND ITS SQUIRTING INSULIN. THE CUSTOMER WAS ADVISED TO DISCONNECT FROM THE PUMP. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 66700 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAS | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723RNAS | A0723RNASJ | 00643169476677 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Other |