FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAS

MDR report key: 7222271 · Received January 26, 2018

Report

Report Number
3004209178-2018-46391
Event Type
Injury
Date Received
January 26, 2018
Date of Event
December 1, 2017
Report Date
May 24, 2018
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
UDI-DI
00643169476677
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE RECEIVED WITH OPERATING CURRENTS WITHIN SPEC. DEVICE PASSED SELF TEST, UNEXPECTED RESTART ERROR TEST, REWIND, BASIC OCCLUSION TEST AND DISPLACEMENT TEST. HOWEVER, UNABLE TO PRIME DEVICE DURING PRIME OR COMPROMISED FORCE SENSOR SYSTEM ALARM TEST DUE TO FAULTY FORCE SENSOR RESISTOR. UNABLE TO PERFORM OCCLUSION TEST, EXCESSIVE NO DELIVERY TEST, DISPLACEMENT ACCURACY TEST OR VERIFY OVER DELIVERY ANOMALY DUE TO PRIME ANOMALY. DEVICE RECEIVED WITH CRACKED BATTERY TUBE THREADS, CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP AND MINOR SCRATCHES ON LCD WINDOW.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT THE CUSTOMER WAS HOSPITALIZED DUE TO DIABETIC KETOACIDOSIS AND HIGH AND LOW BLOOD GLUCOSE TEN YEARS AGO. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 96 MG/DL AND 400 MG/DL. AT TIME OF INCIDENT AND THE BLOOD GLUCOSE LEVEL WAS 96 MG/DL. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 20 MG/DL AND 140-200 MG/DL. THE CUSTOMER WAS TREATED WITH ORANGE JUICE STOPPING TROUBLESHOOT FOR HIGH BLOOD GLUCOSE AND THE CUSTOMER WAS CURRENTLY ON LOW BLOOD GLUCOSE. THE CUSTOMER STATED THAT THE PUMP WAS NOT WORKING FOR THE PAST FEW MONTHS AND ITS SQUIRTING INSULIN. THE CUSTOMER WAS ADVISED TO DISCONNECT FROM THE PUMP. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
66700 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAS PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723RNAS A0723RNASJ 00643169476677

Patients

Seq Age Sex Outcome Treatment
1 71 YR Other