FDA Adverse Event Malfunction Summary report: N

DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 7220574 · Received January 25, 2018

Report

Report Number
3004753838-2018-009369
Event Type
Malfunction
Date Received
January 25, 2018
Date of Event
September 15, 2017
Report Date
February 6, 2018
Manufacturer
DEXCOM, INC.
Product Code
PQF
UDI-DI
10386270000221
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE PATIENT EXPERIENCED ADVERSE EVENTS ON DIFFERENT DAYS WITH THE SAME DEVICE. THE FOLLOWING REPORTS ARE BEING SUBMITTED: 171228-000614.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

DEXCOM WAS MADE AWARE ON (B)(4) 2017 THAT ON (B)(6) 2017, A LOSS OF CONNECTION BETWEEN THE TRANSMITTER AND SMART DEVICE OCCURRED. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. DATA WAS RETURNED AND EVALUATED. THE REPORTED EVENT OF A LOSS OF CONNECTION WAS CONFIRMED VIA DATA. A ROOT CAUSE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

DEXCOM WAS MADE AWARE ON 12/2/2017 THAT ON (B)(6) 2017, A LOSS OF CONNECTION BETWEEN THE TRANSMITTER AND SMART DEVICE OCCURRED. THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THERE WAS NO FAILURE DETECTED RELATED TO THE COMPLAINT. A PAIRING TEST WAS PERFORMED WITH A KNOWN GOOD TRANSMITTER AND PASSED. VOLTAGE TESTING WAS PERFORMED AND PASSED.  A REVIEW OF THE DOWNLOADED DATA LOG CONFIRMED THE REPORTED EVENT OF LOSS OF CONNECTION.  A ROOT CAUSE COULD NOT BE DETERMINED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
62543 DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR PQF DEXCOM, INC. 9438-06 6003906 10386270000221

Patients

Seq Age Sex Outcome Treatment
1 9 YR