CAREASSIST ES120/220/420
Report
- Report Number
- 3006697241-2018-00005
- Event Type
- Malfunction
- Date Received
- January 24, 2018
- Date of Event
- January 11, 2018
- Report Date
- January 11, 2018
- Manufacturer
- HILL-ROM MEXICO
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE HILL-ROM TECHNICIAN FOUND THE RIGHT CAREGIVER CONTROL BOARD NEEDED TO BE REPLACED. PER THE HILL-ROM SERVICE MANUAL THE BED SHOULD BE SUBJECT TO AN EFFECTIVE MAINTENANCE PROGRAM. AN ANNUAL SERVICE OF THE BED IS ADVISED IN ORDER TO MAINTAIN ITS CHARACTERISTICS AND PERFORMANCE. INSPECT THE CABLE ROUTING FOR PINCHING, BINDING, AND DAMAGE. MAKE SURE ALL FUNCTIONS ON THE CAREGIVER CONTROL OPERATE CORRECTLY. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE RIGHT CAREGIVER CONTROL BOARD TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING THE HILOW GOES UP BY ITSELF. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).