FDA Adverse Event Malfunction Summary report: N

BD PRECISIONGLIDE¿ NEEDLE

MDR report key: 7212523 · Received January 23, 2018

Report

Report Number
1911916-2018-00006
Event Type
Malfunction
Date Received
January 23, 2018
Date of Event
January 2, 2018
Report Date
February 27, 2018
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMI
UDI-DI
30382903051961
PMA / PMN Number
K021475
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: QN REVIEW: NO NOTIFICATIONS WERE WRITTEN FOR THIS BATCH, NOR FOR THE ASSOCIATED ASSEMBLY BATCH. DHR REVIEW: ASSEMBLY BATCH 7060534 HAD (B)(4) VISUAL INSPECTIONS PERFORMED ON (B)(4) PARTS WITH ONE DEFECT NOTED FOR INSUFFICIENT EPOXY. ALL AFFECTED PRODUCT WAS SCRAPPED IN-PROCESS. INVESTIGATION RESULTS: NOT APPLICABLE; NO SAMPLE OR PHOTO WERE PROVIDED. BASED ON THE DESCRIPTION, THE FM IS PROBABLY AN EPOXY SPLATTER ON THE CANNULA. POSSIBLE ROOT CAUSE: PROBABLE ROOT CAUSES: THE RACK OF NEEDLES FELL AGAINST ANOTHER RACK COMING IN CONTACT WITH UNCURED EPOXY. (MOST LIKELY IN VIEW OF THE NOTE IN THE DHR REGARDING A ROLL UP.) A MISASSEMBLED NEEDLE WITH UNCURED EPOXY CAME INTO CONTACT WITH THE NEEDLES IN QUESTION DURING ASSEMBLY. EPOXY CONTAMINATED THE CATERPILLAR BELTS ON THE ASSEMBLY LINE AND TRANSFERRED TO THE CANNULA.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BEFORE USE A BD PRECISIONGLIDE¿ NEEDLE WAS FOUND WITH ¿A WHITE SPOT ON THE NEEDLE WHEN OPENED IN THE CACI. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
55211 BD PRECISIONGLIDE¿ NEEDLE SAFETY ENGINEERED NEEDLE FMI BD MEDICAL (BD WEST) MEDICAL SURGICAL 7086929 30382903051961

Patients

Seq Age Sex Outcome Treatment
1 Other