FDA Adverse Event Injury Summary report: N

MITRACLIP SYSTEM CLIP DELIVERY SYSTEM

MDR report key: 7211247 · Received January 22, 2018

Report

Report Number
2024168-2018-00519
Event Type
Injury
Date Received
January 22, 2018
Date of Event
September 29, 2017
Report Date
February 13, 2018
Manufacturer
AV-TEMECULA-CT
Product Code
NKM
PMA / PMN Number
P100009
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Description of Event or Problem · 0

Unique Complaint ID Number,Initial or Supplement,Type of event,TTE in days,Device Brand Name,Medical device identifier,Device Codes;336527?1?1 ?;3102329,,SI,926,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3102581,,SI,412,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3102581,,SI,739,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3102581,,SI,838,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3102799,,SI,780,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3102799,,SI,825,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3103831,,SI,608,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3103831,,SI,857,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3103831,,SI,876,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3104555,,SI,646,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3104555,,SI,808,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3104717,,SI,913,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3104734,,SI,505,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3104734,,SI,690,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3104734,,SI,690,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3106543,,SI,905,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3106937,,SI,369,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3107854,,SI,485,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3150617,,SI,714,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3150617,,SI,724,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3150617,,SI,831,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3150617,,SI,850,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3152635,,SI,698,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3155284,,SI,833,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3157132,,SI,369,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3161538,,SI,400,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3161871,,SI,574,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3164808,,SI,528,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3165090,,SI,504,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3165090,,SI,513,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3165090,,SI,513,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3167163,,SI,595,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3167163,,SI,604,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3168362,,SI,378,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3168362,,SI,409,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3168362,,SI,648,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3168362,,SI,904,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3168362,,SI,911,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3168362,,SI,378,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3169109,,SI,898,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3169109,,SI,923,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3170688,,SI,513,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3170688,,SI,513,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3173016,,SI,502,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3173016,,SI,507,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3173016,,SI,528,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3173016,,SI,528,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3173043,,SI,528,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3173043,,SI,528,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3173273,,SI,745,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3174028,,SI,398,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3174143,,SI,515,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3175440,,SI,434,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3175440,,SI,435,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3175440,,SI,875,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3178548,,SI,459,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3179217,,SI,643,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3179217,,SI,643,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3180093,,SI,823,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3181823,,SI,552,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3181874,,SI,472,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3181874,,SI,580,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3181874,,SI,580,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3185355,,SI,604,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3186340,,SI,651,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3186340,,SI,679,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3186340,,SI,651,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3187061,,SI,453,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3187061,,SI,472,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3187061,,SI,532,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3187061,,SI,533,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3187061,,SI,548,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3187061,,SI,659,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3187061,,SI,453,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3187061,,SI,659,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3187749,,SI,360,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3187749,,SI,680,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3188965,,SI,511,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3188965,,SI,511,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3189108,,SI,339,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3189108,,SI,344,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3189108,,SI,339,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3191198,,SI,574,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3191342,,SI,389,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3191730,,SI,691,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3191730,,SI,767,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3195601,,SI,717,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3196592,,SI,838,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3196592,,SI,867,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3199006,,SI,650,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3199058,,SI,725,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3199058,,SI,716,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3201855,,SI,426,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3201855,,SI,434,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3201855,,SI,821,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3201855,,SI,894,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3202391,,SI,500,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3202845,,SI,612,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3202845,,SI,612,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3204814,,SI,799,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3204886,,SI,544,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3204886,,SI,636,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3204886,,SI,705,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3204886,,SI,544,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3205447,,SI,636,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3205447,,SI,635,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3205447,,SI,636,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3207935,,SI,694,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3252456,,SI,520,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3253664,,SI,423,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3253664,,SI,586,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3254319,,SI,611,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3254319,,SI,617,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3254319,,SI,680,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3254319,,SI,439,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3255013,,SI,609,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3255013,,SI,612,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3255806,,SI,454,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3255822,,SI,407,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3304401,,SI,515,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3304401,,SI,729,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3304401,,SI,798,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3304401,,SI,515,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3306456,,SI,555,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3306456,,SI,555,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3306551,,SI,502,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3306551,,SI,511,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3306551,,SI,537,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3306551,,SI,573,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3306551,,SI,681,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3306551,,SI,682,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3307059,,SI,482,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3307102,,SI,749,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3307102,,SI,749,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3307251,,SI,446,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3310505,,SI,501,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3310505,,SI,536,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3310505,,SI,542,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3310761,,SI,378,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3311774,,SI,528,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3311928,,SI,640,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3311928,,SI,640,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3312671,,SI,512,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3312671,,SI,526,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3312753,,SI,700,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3312940,,SI,651,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3312940,,SI,651,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3314114,,SI,495,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3314223,,SI,842,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3314543,,SI,569,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3314543,,SI,573,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3314543,,SI,589,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3314973,,SI,515,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3314973,,SI,515,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3318560,,SI,392,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3318560,,SI,407,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3318560,,SI,413,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3318560,,SI,407,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3318560,,SI,413,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3319265,,SI,462,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3319654,,SI,460,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3319654,,SI,533,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3319654,,SI,717,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3319714,,SI,519,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3320869,,SI,403,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3322868,,SI,483,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3322868,,SI,494,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3322868,,SI,543,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3322868,,SI,682,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3322868,,SI,543,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3324937,,SI,527,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3324937,,SI,649,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3324937,,SI,740,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3324937,,SI,800,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3324937,,SI,800,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3324999,,SI,531,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3325123,,SI,422,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3325123,,SI,472,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3325123,,SI,557,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3325123,,SI,694,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3326958,,SI,768,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3327231,,SI,784,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3327231,,SI,784,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3328596,,SI,396,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3328596,,SI,441,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3328596,,SI,441,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3329181,,SI,808,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3329326,,SI,388,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3329326,,SI,446,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3329891,,SI,764,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3330727,,SI,652,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3330782,,SI,679,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3330782,,SI,696,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3330782,,SI,776,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3330782,,SI,696,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3332998,,SI,764,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3333627,,SI,502,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3335676,,SI,622,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3335676,,SI,649,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3335815,,SI,437,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3336930,,SI,386,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3336930,,SI,423,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3336930,,SI,386,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3338750,,SI,531,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3361803,,SI,432,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3361803,,SI,716,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3362375,,SI,476,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3362375,,SI,531,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3362587,,SI,344,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3362587,,SI,435,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3362587,,SI,637,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3364190,,SI,496,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3364190,,SI,688,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3364190,,SI,688,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3368691,,SI,413,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3371439,,SI,608,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3371439,,SI,627,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3371439,,SI,608,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3373255,,SI,581,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3373335,,SI,594,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3374045,,SI,373,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3374175,,SI,552,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3374175,,SI,552,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3378667,,SI,594,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3379368,,SI,463,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3381507,,SI,761,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3381702,,SI,420,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3383212,,SI,479,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3383212,,SI,513,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3383235,,SI,403,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3384736,,SI,1000,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3386408,,SI,633,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3386408,,SI,659,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3386408,,SI,659,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3388094,,SI,366,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3388450,,SI,974,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3388450,,SI,974,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3388546,,SI,562,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3389564,,SI,639,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3391937,,SI,928,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3395270,,SI,693,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3397147,,SI,390,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3397147,,SI,405,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3397147,,SI,503,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3397147,,SI,390,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3397147,,SI,503,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3397389,,SI,579,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3397500,,SI,549,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3397500,,SI,732,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3398485,,SI,392,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3398485,,SI,460,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3398485,,SI,487,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3398485,,SI,526,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3398485,,SI,487,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3398485,,SI,502,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3398485,,SI,526,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3399287,,SI,470,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3399287,,SI,578,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3399287,,SI,470,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3400220,,SI,383,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3400220,,SI,430,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3400220,,SI,430,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3400629,,SI,565,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3400629,,SI,568,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3400629,,SI,590,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3404409,,SI,446,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3404409,,SI,588,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3405049,,SI,588,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3405049,,SI,588,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3406976,,SI,405,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3406976,,SI,483,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3406976,,SI,590,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3406976,,SI,405,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3406976,,SI,483,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3411283,,SI,422,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3411283,,SI,540,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3411283,,SI,540,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3411283,,SI,422,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3412306,,SI,560,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3412306,,SI,569,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3412306,,SI,651,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3415738,,SI,417,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3415757,,SI,596,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3415757,,SI,665,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3415757,,SI,596,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3416024,,SI,705,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3416078,,SI,552,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3416458,,SI,493,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3441229,,SI,595,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3482887,,SI,476,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3483153,,SI,366,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3483153,,SI,393,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3483153,,SI,444,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3483153,,SI,488,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3483153,,SI,498,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3483153,,SI,506,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3483153,,SI,515,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3483153,,SI,366,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3483153,,SI,393,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3483153,,SI,444,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3483153,,SI,488,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3483664,,SI,387,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3483712,,SI,621,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3483712,,SI,621,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3485190,,SI,476,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3485190,,SI,513,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3485190,,SI,535,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3487614,,SI,625,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3488604,,SI,445,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3489131,,SI,427,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3490547,,SI,595,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3490547,,SI,595,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3490640,,SI,592,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3490640,,SI,601,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3490640,,SI,606,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3493209,,SI,733,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3493576,,SI,484,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3493576,,SI,630,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3493576,,SI,649,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3493576,,SI,654,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3493587,,SI,400,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3493844,,SI,583,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3494796,,SI,468,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3494833,,SI,462,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3494833,,SI,469,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3494833,,SI,469,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3497835,,SI,496,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3499580,,SI,497,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3499580,,SI,499,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3500985,,SI,439,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3500985,,SI,439,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3594268,,SI,653,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3594268,,SI,659,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078905,,SI,809,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078905,,SI,546,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078928,,SI,720,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078928,,SI,788,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078928,,SI,642,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078928,,SI,720,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078928,,SI,788,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078963,,SI,965,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3079293,,SI,372,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3079293,,SI,440,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3079293,,SI,955,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3079293,,SI,1052,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3079293,,SI,1066,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3082606,,SI,651,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3082606,,SI,724,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3082606,,SI,761,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3083654,,SI,928,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3083784,,SI,448,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3083784,,SI,468,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3083784,,SI,468,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3083819,,SI,616,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3083819,,SI,636,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3084039,,SI,728,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3084885,,SI,386,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3084885,,SI,1022,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3084885,,SI,1074,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3084954,,SI,376,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3084954,,SI,399,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3084954,,SI,465,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3084954,,SI,472,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3084954,,SI,533,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3084954,,SI,762,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3084954,,SI,908,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3084954,,SI,965,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3084954,,SI,1000,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3084954,,SI,1146,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3084954,,SI,762,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3084954,,SI,908,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3084954,,SI,965,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3107854,,SI,498,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3108163,,SI,363,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3108163,,SI,366,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3108163,,SI,453,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3108163,,SI,751,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3108393,,SI,764,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3108393,,SI,764,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3109350,,SI,758,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3109350,,SI,765,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3109350,,SI,765,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3109666,,SI,981,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3109666,,SI,981,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;1241191,,SI,414,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;1241191,,SI,432,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;1241191,,SI,432,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;1247778,,SI,617,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;1247778,,SI,642,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;1247778,,SI,770,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;1268082,,SI,402,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;1268082,,SI,676,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;1283536,,SI,778,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;1670604,,SI,518,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;1670604,,SI,628,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;1947385,,SI,615,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;1947385,,SI,705,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;1947385,,SI,728,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;1947385,,SI,561,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;1947385,,SI,571,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;1947385,,SI,705,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2355254,,SI,934,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2653003,,SI,973,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2747667,,SI,718,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2747667,,SI,746,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2747667,,SI,762,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2747667,,SI,718,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2747667,,SI,746,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2793238,,SI,639,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2871169,,SI,492,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2878425,,SI,844,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2878425,,SI,844,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2961640,,SI,376,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2961640,,SI,401,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2963665,,SI,376,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2963665,,SI,471,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2963665,,SI,522,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2963665,,SI,571,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2963665,,SI,581,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2963665,,SI,471,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2963665,,SI,522,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2982765,,SI,511,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2982765,,SI,694,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2983145,,SI,877,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2990557,,SI,491,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;2990557,,SI,626,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3032176,,SI,520,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3032176,,SI,929,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3041308,,SI,643,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3041308,,SI,684,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3043765,,SI,499,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3043765,,SI,677,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3043765,,SI,744,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3043765,,SI,793,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3074641,,SI,420,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3074641,,SI,423,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3075060,,SI,608,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3075060,,SI,693,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3075060,,SI,693,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3075180,,SI,850,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3075180,,SI,949,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3075180,,SI,1036,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3075681,,SI,599,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3075681,,SI,863,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3076771,,SI,697,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3077576,,SI,383,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3077576,,SI,406,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3077576,,SI,412,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3077576,,SI,446,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3077576,,SI,383,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3077576,,SI,406,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3077697,,SI,499,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3077697,,SI,666,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3077713,,SI,605,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3077713,,SI,624,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3077713,,SI,719,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078699,,SI,534,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078699,,SI,852,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078699,,SI,876,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078699,,SI,852,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078851,,SI,975,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078851,,SI,1036,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078851,,SI,1066,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078851,,SI,1079,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078891,,SI,406,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078891,,SI,707,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078891,,SI,737,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078891,,SI,752,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078891,,SI,785,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078891,,SI,1071,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078891,,SI,1073,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078891,,SI,752,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078891,,SI,785,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078891,,SI,1071,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078891,,SI,1073,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078892,,SI,457,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078892,,SI,465,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078892,,SI,503,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078905,,SI,415,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078905,,SI,546,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078905,,SI,690,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078905,,SI,721,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078905,,SI,744,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3078905,,SI,765,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3147872,,SI,600,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3150617,,SI,412,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3207935,,SI,716,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3207935,,SI,716,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3208414,,SI,473,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3208414,,SI,689,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3208414,,SI,728,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3208414,,SI,689,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3209916,,SI,524,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3209916,,SI,822,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3209916,,SI,833,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3210413,,SI,720,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3210413,,SI,879,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3210651,,SI,814,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3210651,,SI,814,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3211149,,SI,613,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3211378,,SI,594,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3211378,,SI,594,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3211662,,SI,642,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3211662,,SI,658,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3211752,,SI,498,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3212379,,SI,654,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3212379,,SI,700,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3212379,,SI,779,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3212379,,SI,779,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3215401,,SI,581,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3215828,,SI,440,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3215828,,SI,541,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3218087,,SI,535,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3218087,,SI,551,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3218087,,SI,631,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3218087,,SI,720,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3218087,,SI,731,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3218087,,SI,738,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3218087,,SI,758,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3218087,,SI,767,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3218087,,SI,551,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3218087,,SI,731,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3218087,,SI,738,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3218087,,SI,758,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3219677,,SI,758,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3222062,,SI,472,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3222062,,SI,716,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3222320,,SI,851,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3222320,,SI,968,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3222320,,SI,975,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3222320,,SI,851,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3222320,,SI,968,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3222320,,SI,975,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3223212,,SI,673,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3223361,,SI,521,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3224519,,SI,527,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3224519,,SI,549,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3224519,,SI,591,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3224519,,SI,596,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3224519,,SI,527,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3224519,,SI,596,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3226026,,SI,568,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3226336,,SI,761,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3226336,,SI,765,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3226336,,SI,770,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3226336,,SI,798,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3226336,,SI,761,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3226336,,SI,798,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3227493,,SI,589,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3227522,,SI,625,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3228494,,SI,500,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3228494,,SI,681,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3228494,,SI,716,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3228494,,SI,500,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3228494,,SI,681,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3228823,,SI,449,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3228823,,SI,525,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3228823,,SI,533,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3228944,,SI,555,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3229475,,SI,429,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3229475,,SI,578,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3229475,,SI,628,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3229475,,SI,631,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3229475,,SI,670,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3229475,,SI,429,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3230759,,SI,373,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3230759,,SI,373,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3231838,,SI,520,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3231838,,SI,577,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3231838,,SI,624,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3231838,,SI,732,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3231838,,SI,776,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3231838,,SI,865,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3232169,,SI,568,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3233338,,SI,637,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3233338,,SI,760,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3233602,,SI,516,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3234539,,SI,989,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3234648,,SI,787,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3237563,,SI,849,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3237689,,SI,602,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3237819,,SI,480,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3237819,,SI,598,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3237819,,SI,819,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3237819,,SI,480,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3237862,,SI,778,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3237862,,SI,792,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3241176,,SI,428,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3241176,,SI,458,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3243344,,SI,482,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3243900,,SI,403,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3243900,,SI,423,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3243900,,SI,423,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3247243,,SI,401,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3339357,,SI,682,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3344248,,SI,489,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3344248,,SI,522,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3344886,,SI,427,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3344886,,SI,453,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3344886,,SI,547,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3344886,,SI,594,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3344886,,SI,736,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3344886,,SI,747,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3344886,,SI,427,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3344886,,SI,453,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3344886,,SI,547,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3348284,,SI,747,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3350019,,SI,406,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3350019,,SI,628,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3350019,,SI,639,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3350019,,SI,671,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3350030,,SI,520,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3350030,,SI,560,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3351504,,SI,728,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3352322,,SI,495,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3352330,,SI,374,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3354911,,SI,370,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3354911,,SI,395,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3355213,,SI,695,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3355620,,SI,673,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3355648,,SI,605,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3356249,,SI,584,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3358102,,SI,497,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3358249,,SI,675,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3358873,,SI,661,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3359284,,SI,376,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3359284,,SI,418,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3359284,,SI,376,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3361803,,SI,432,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3361803,,SI,716,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3416458,,SI,586,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3416458,,SI,586,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3422036,,SI,554,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3423123,,SI,380,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3423123,,SI,457,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3424078,,SI,374,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3424078,,SI,406,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3424078,,SI,617,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3424078,,SI,696,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3424078,,SI,701,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3424078,,SI,696,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3424689,,SI,405,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3425380,,SI,476,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3425380,,SI,626,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3426056,,SI,765,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3426056,,SI,772,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3426056,,SI,773,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3426056,,SI,765,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3427413,,SI,420,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3427413,,SI,436,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3427413,,SI,492,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3427413,,SI,513,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3427413,,SI,526,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3427413,,SI,557,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3427413,,SI,600,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3427413,,SI,618,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3427413,,SI,652,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3427413,,SI,675,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3427413,,SI,492,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3427413,,SI,526,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3429005,,SI,560,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3429005,,SI,560,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3429924,,SI,362,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3429924,,SI,553,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3431135,,SI,637,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3432685,,SI,627,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3434265,,SI,627,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3434265,,SI,627,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3434342,,SI,410,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3434583,,SI,395,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3434583,,SI,399,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3438855,,SI,562,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3438855,,SI,664,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3438855,,SI,740,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3438855,,SI,562,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3439978,,SI,534,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3439978,,SI,534,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3440917,,SI,707,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3084954,,SI,1000,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3085234,,SI,664,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3085234,,SI,666,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3085234,,SI,692,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3085234,,SI,692,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3085677,,SI,614,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3085677,,SI,805,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3085677,,SI,882,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3085677,,SI,885,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3085677,,SI,805,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3085832,,SI,736,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3086259,,SI,405,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3086412,,SI,453,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3086412,,SI,883,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3086412,,SI,1003,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3086412,,SI,883,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3086888,,SI,554,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3086888,,SI,800,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3086888,,SI,800,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3086923,,SI,851,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3086923,,SI,862,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3086923,,SI,943,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3086923,,SI,862,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3088362,,SI,595,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3088394,,SI,548,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3088394,,SI,548,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3088502,,SI,815,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3088643,,SI,677,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3088643,,SI,715,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3090787,,SI,420,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3091068,,SI,504,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3091628,,SI,510,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3091628,,SI,523,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3091628,,SI,566,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3091628,,SI,1061,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3091628,,SI,1082,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3091628,,SI,566,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3092409,,SI,378,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3092409,,SI,471,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3092517,,SI,813,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3093135,,SI,391,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3093187,,SI,385,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3093522,,SI,412,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3093522,,SI,643,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3093522,,SI,390,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3093773,,SI,847,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3094412,,SI,725,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3094412,,SI,769,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3094412,,SI,852,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3094412,,SI,452,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3094412,,SI,494,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3094914,,SI,682,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3094914,,SI,706,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3095287,,SI,1000,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3095287,,SI,1016,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3095362,,SI,525,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3095362,,SI,535,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3095362,,SI,646,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3095362,,SI,1077,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3095811,,SI,658,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3095811,,SI,795,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3095811,,SI,846,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3095811,,SI,846,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3095827,,SI,941,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3095827,,SI,941,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3096191,,SI,940,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3096191,,SI,981,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3096191,,SI,423,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3096349,,SI,799,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3096349,,SI,799,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3096412,,SI,844,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3096412,,SI,844,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3097009,,SI,717,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3097009,,SI,743,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3097028,,SI,638,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3097252,,SI,528,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3097673,,SI,371,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3097673,,SI,883,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3097735,,SI,678,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3097819,,SI,729,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3097819,,SI,855,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3097819,,SI,876,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3097819,,SI,892,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3097819,,SI,921,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3097819,,SI,892,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3097819,,SI,855,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3099104,,SI,497,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3099104,,SI,740,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3099104,,SI,1055,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3099955,,SI,596,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3099955,,SI,965,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3100540,,SI,373,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3100540,,SI,728,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3101382,,SI,590,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3101555,,SI,754,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3102047,,SI,388,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3102047,,SI,388,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3102176,,SI,704,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3102329,,SI,912,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3111014,,SI,707,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3111014,,SI,802,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3111014,,SI,879,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3111209,,SI,601,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3111298,,SI,565,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3111298,,SI,655,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3111298,,SI,895,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3111538,,SI,682,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3111538,,SI,991,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3111538,,SI,682,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3111938,,SI,881,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3112403,,SI,742,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3112403,,SI,979,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3112765,,SI,714,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3112796,,SI,491,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3113533,,SI,786,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3114340,,SI,789,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3114340,,SI,926,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3117202,,SI,371,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3117202,,SI,442,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3117202,,SI,481,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3117202,,SI,597,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3117202,,SI,618,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3117202,,SI,666,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3117202,,SI,764,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3117202,,SI,785,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3118427,,SI,945,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3118427,,SI,954,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3118427,,SI,862,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3118427,,SI,924,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3123411,,SI,917,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3123411,,SI,634,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3128144,,SI,668,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3128402,,SI,895,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3131442,,SI,657,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3131442,,SI,859,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3131442,,SI,859,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3131517,,SI,728,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3132570,,SI,399,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3132570,,SI,486,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3132570,,SI,759,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3132570,,SI,842,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3132570,,SI,1059,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3135401,,SI,378,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3135401,,SI,398,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3135401,,SI,420,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3135401,,SI,422,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3135401,,SI,398,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3137912,,SI,609,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3137912,,SI,620,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3138062,,SI,793,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3138062,,SI,793,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3139361,,SI,406,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3140165,,SI,880,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3140165,,SI,892,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3140165,,SI,774,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3141920,,SI,407,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3147872,,SI,600,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3255822,,SI,501,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3255822,,SI,564,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3255822,,SI,576,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3256878,,SI,410,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3256878,,SI,759,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3259789,,SI,475,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3260124,,SI,764,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3260805,,SI,772,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3260805,,SI,772,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3261514,,SI,708,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3261514,,SI,708,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3265287,,SI,708,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3265618,,SI,518,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3265618,,SI,518,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3266506,,SI,622,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3266870,,SI,462,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3266870,,SI,659,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3266870,,SI,750,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3266870,,SI,750,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3268503,,SI,417,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3268503,,SI,520,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3268503,,SI,773,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3269257,,SI,596,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3269257,,SI,596,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3270136,,SI,533,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3271337,,SI,586,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3271337,,SI,590,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3273573,,SI,430,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3273573,,SI,436,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3273573,,SI,474,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3273573,,SI,604,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3273573,,SI,436,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3273573,,SI,474,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3273573,,SI,604,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3274551,,SI,555,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3274551,,SI,595,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3276281,,SI,409,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3276360,,SI,559,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3276360,,SI,665,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3276360,,SI,449,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3276360,,SI,559,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3276704,,SI,499,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3276704,,SI,594,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3276704,,SI,499,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3276704,,SI,594,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3277659,,SI,535,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3279348,,SI,391,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3279348,,SI,402,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3279348,,SI,611,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3279467,,SI,671,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3279467,,SI,707,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3279467,,SI,671,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3280006,,SI,758,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3282454,,SI,568,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3282454,,SI,595,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3282454,,SI,597,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3282454,,SI,608,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3283167,,SI,397,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3286433,,SI,446,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3286433,,SI,476,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3286433,,SI,647,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3287906,,SI,413,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3289220,,SI,819,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3289220,,SI,819,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3290288,,SI,745,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3291090,,SI,432,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3291090,,SI,566,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3291090,,SI,610,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3291090,,SI,566,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3292453,,SI,754,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3293455,,SI,519,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3293455,,SI,519,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3294887,,SI,368,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3295434,,SI,373,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3295434,,SI,373,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3297022,,SI,681,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3297022,,SI,728,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3297022,,SI,744,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3297022,,SI,728,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3299119,,SI,429,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3299119,,SI,697,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3302144,,SI,642,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3302144,,SI,703,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3302144,,SI,758,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3302144,,SI,763,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3302144,,SI,793,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3302144,,SI,813,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3303891,,SI,497,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3391937,,SI,928,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3392246,,SI,477,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3392246,,SI,500,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3393075,,SI,447,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3393775,,SI,667,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3393775,,SI,667,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3394670,,SI,475,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3394670,,SI,475,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3394845,,SI,705,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3394845,,SI,775,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3441270,,SI,482,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3441270,,SI,496,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3441270,,SI,525,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3441270,,SI,496,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3441438,,SI,517,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3441438,,SI,517,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3442011,,SI,443,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3442011,,SI,619,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3442276,,SI,532,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3442276,,SI,532,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3442558,,SI,432,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3442574,,SI,658,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3444854,,SI,566,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3444854,,SI,652,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3444854,,SI,652,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3445222,,SI,502,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3445646,,SI,523,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3446008,,SI,580,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3449013,,SI,415,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3449013,,SI,486,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3449111,,SI,511,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3449111,,SI,496,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3452424,,SI,437,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3452424,,SI,473,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3452424,,SI,572,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3452424,,SI,473,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3453950,,SI,400,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3453950,,SI,433,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3454032,,SI,429,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3454032,,SI,429,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3454413,,SI,559,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3457406,,SI,551,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3457406,,SI,551,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3457437,,SI,493,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3457437,,SI,526,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3457437,,SI,557,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3457437,,SI,557,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3461125,,SI,382,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3461125,,SI,575,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3461428,,SI,494,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3461428,,SI,653,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3463455,,SI,463,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3464898,,SI,610,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3464898,,SI,610,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3464923,,SI,492,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3465146,,SI,463,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3465410,,SI,634,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3465410,,SI,647,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3466782,,SI,556,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3466782,,SI,569,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3467675,,SI,508,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3470561,,SI,383,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3470561,,SI,399,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3470561,,SI,408,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3470561,,SI,450,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3470561,,SI,383,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3470561,,SI,399,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3471561,,SI,395,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3472571,,SI,379,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3472571,,SI,499,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3472571,,SI,516,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3472571,,SI,527,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3472571,,SI,546,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3472571,,SI,583,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3472571,,SI,612,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3474568,,SI,470,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3474568,,SI,470,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3474933,,SI,595,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3476284,,SI,499,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3476806,,SI,626,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3476806,,SI,631,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3479184,,SI,407,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3479184,,SI,407,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3479232,,SI,631,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3481517,,SI,572,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3481517,,SI,577,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3482376,,SI,392,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3482376,,SI,431,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3482376,,SI,392,MITRACLIP SYSTEM Clip Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;1947385,,SI,571,MITRACLIP SYSTEM Clip;Delivery System,N/A,2993 ? No Known Device Problem;336527?1?1 ?;3474310,,SI,421,MITRACLIP SYSTEM Clip;Delivery System,N/A,2993 ? No Known Device Problem

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. DATE OF EVENT, DATE OF IMPLANT - ESTIMATED. DUE TO SYSTEM LIMITATIONS, THE TYPE OF REPORT IS SELECTED AS 30-DAY ALTHOUGH THIS IS A PERIODIC REPORT. PAS1 AND PAS2 MITRACLIP POST APPROVAL STUDY DATA IS REPORTED AS A SUMMARY PER SUMMARY REPORTING EXCEPTION APPROVAL NUMBER E2015009. THE DEVICES ARE NOT RETURNING. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS NOT RETURNED FOR ANALYSIS. ADDITIONALLY, A REVIEW OF THE LOT HISTORY RECORDS COULD NOT BE CONDUCTED, BECAUSE THE LOT NUMBERS WERE NOT PROVIDED. THE REPORTED PATIENT EFFECTS OF ATRIAL FIBRILLATION, MYOCARDIAL INFARCTION, TRANSIENT ISCHEMIC ATTACK (TIA) AND WORSENING HEART FAILURE, ARE LISTED IN THE MITRACLIP SYSTEM ELECTRONIC INSTRUCTIONS FOR USE AS KNOWN POSSIBLE COMPLICATIONS ASSOCIATED WITH MITRACLIP PROCEDURES. BASED ON THE INFORMATION REVIEWED, A CONCLUSIVE CAUSE FOR THE REPORTED PATIENT EFFECTS, AND THE RELATIONSHIP TO THE DEVICE, IF ANY, CANNOT BE DETERMINED. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN, OR LABELING.

Description of Event or Problem · 1

PER THE REQUIREMENTS OF THE POST APPROVAL STUDY 1 (PAS1) AND PAS2, POST APPROVAL MITRACLIP STUDIES CONDUCTED VIA THE TVT REGISTRY, ABBOTT HAS RECEIVED DATA DOWNLOADED FROM THE CENTER FOR MEDICARE AND MEDICAID SERVICES (CMS) CLAIMS DATABASE FOR PATIENTS IN THE STUDIES. THE DATA CONTAINED 1014 ADVERSE EVENTS. THE RELATIONSHIP OF THE EVENTS TO THE MITRACLIP DEVICE CANNOT BE DETERMINED BASED ON THE DATA RECEIVED; THEREFORE, ABBOTT VASCULAR IS CONSERVATIVELY REPORTING THESE EVENTS. THE DATA IS REPORTED AS A SUMMARY REPORT PER SUMMARY REPORTING EXEMPTION APPROVAL NUMBER E2015009.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
52386 MITRACLIP SYSTEM CLIP DELIVERY SYSTEM MITRACLIP DELIVERY SYSTEM NKM AV-TEMECULA-CT

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R| S